RecallWatchUSA

Home / Recalls / Medical Devices

Biomet, Inc. recalls BIOMET Trauma, [Item # 13182-22-26, DVR CROSSLOCK EXTRA LONG PLATE LEFT], [Item # 1318-12-126, DVR CROSSLOCK EXTRA LONG

Recalled March 30, 2022 · FDA recall Z-0810-2022 · Biomet

Hazard
Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part. · FDA Class II
Units
38
Sold at
Worldwide: US (28 States AL, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MD, MI, MS, NC, NE, NM, NY, OH, OK, OR, SC, TX, UT, WA, WY), Australia, Japan, and Netherlands

Description

Packaging of various implants may not have sufficient adhesion on all sides, leading to loss of sterile barrier integrity. Use of product poses risk of infection, leading to surgical intervention, and/or extension of surgery to find another part.

Product
BIOMET Trauma, [Item # 13182-22-26, DVR CROSSLOCK EXTRA LONG PLATE LEFT], [Item # 1318-12-126, DVR CROSSLOCK EXTRA LONG PLATE RIGHT], [Item # 1318-12-176, DVR CROSSLOCK EXTRA EXTRA LONG PLATE RIGHT]. Devices are packaged in a vacuum sealed nylon pouch, which is placed in a carton with the appropriate IFU and patient record labels. Identifying labels on the pouch and the carton.
Manufacturer(s)
Biomet, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov