RecallWatchUSA

Home / Recalls / Medical Devices

BioMerieux SA recalls VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments

Recalled April 19, 2017 · FDA recall Z-1801-2017 · BioMerieux SA

Hazard
Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant. · FDA Class II
Units
834 kits
Sold at
Nationwide Distribution including AZ, CA, CO, FL, GA, IN, IA, LA, MI, MS, NY, NC, OH, OK, OR, PA, TX, WA

Description

Fulvestrant cross reacts with the VIDAS Estradiol II immunoassay and may cause falsely elevated estradiol results in patients treated with Fulvestrant.

Product
VIDAS Estradiol II, Ref 30431-01, 60 tests The VIDAS Estradiol II (E2 II) assay is intended for use on the instruments of the VIDAS family (VITEK ImmunoDiagnostic Assay System) as an automated quantitative enzyme-linked fluorescent immunoassay (ELFA) to the determination of total estradiol concentration in human serum or plasma (heparin).
Manufacturer(s)
BioMerieux SA
Made in
France

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov