RecallWatchUSA

Home / Recalls / Medical Devices

Becton Dickinson & Company recalls SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 008

Recalled July 29, 2020 · FDA recall Z-2669-2020 · Becton Dickinson &

Hazard
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection. · FDA Class II
Units
5,296,132 total
Sold at
Worldwide Distribution

Description

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

Product
SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 00801741066177), LH-0030YN (UDI 00801741066191), LH-0031YN (UDI 00801741066214), LH-0032YN (UDI 00801741066238), LH-0033YN (UDI 00801741066252), LH-0034YN (UDI 00801741066276), LH-0035YN (UDI 00801741066290 ), LH-0036YN (UDI 00801741066313), LH-0037YN (UDI 00801741066337), LH-0038YN (UDI 00801741066351), LH-0039YN (UDI 00801741066313), LH-0040YN (UDI 00801741066399)
Manufacturer(s)
Becton Dickinson & Company

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Becton Dickinson & Company recalls SafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog Numbers LH-0029YN (UDI 008 | RecallWatchUSA