RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter, Inc. recalls UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended

Recalled December 4, 2019 · FDA recall Z-0516-2020 · Beckman Coulter

Hazard
When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment. · FDA Class II
Units
1,056 (418 US, and 638 OUS)
Sold at
Worldwide distribution - US Nationwide distribution and countries of Australia, Bahrain, Belarus, Belgium, Brazil, Canada, China, Colombia, Czech Republic, Egypt, Ethiopia, France, Germany, Greece, Hong Kong, India, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Macau, Malaysia, Mayotte, Mexico, Morocco, Netherlands, New Zealand, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Vietnam.

Description

When the UniCel DxH Slidemaker Stainer and UniCel DxH Slidemaker Stainer II dispense probe is slightly bent, there is a potential for cell carryover from the previous sample, and erroneous results could be reported from the slide review. Erroneous results may be reported from the laboratory which may trigger unnecessary or inappropriate patient treatment.

Product
UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended use is as follows: The DxH Slidemaker Stainer is a fully automated slide preparation and staining device that aspirates a whole-blood sample, smears a blood film on a clean microscope slide, and delivers a variety of fixatives, stains, buffers, and rinse solutions to that blood smear.
Manufacturer(s)
Beckman Coulter, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Beckman Coulter, Inc. recalls UniCel DxH Slidemaker Stainer (DxH SMS) and UniCel DxH Slidemaker Stainer II (DxH SMS II) - Product Usage: The intended | RecallWatchUSA