Beckman Coulter Inc. recalls UniCel¿ DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel¿ DxC Synchron¿ Clinical Systems are fully

Recalled October 31, 2012 · FDA recall Z-0123-2013 · Beckman Coulter

Hazard
Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the Shuttle to Bar · FDA Class II
Units
7 units total (both US and International)
Sold at
Worldwide Distribution - USA including TX and OH. Interntionally to China and Turkey

Description

Beckman Coulter has confirmed that the Shuttle to Barcode Reader and Cap Piercer alignment procedure cannot be performed on certain DxC configurations with v5.0.11 software installed. If the Shuttle to Barcode Reader becomes misaligned, the Operator will not be able to perform this alignment procedure on the DxC systems listed above. Software v5.0 .11 does not allow access to the Shuttle to Bar

Product
UniCel¿ DxC 600/800 with Software v5.0.11, Part Number B12218. The UniCel¿ DxC Synchron¿ Clinical Systems are fully automated, computercontrolled clinical chemistry analyzers designed for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries. Analysis can be performed on serum, plasma, urine, or cerebrospinal fluid (CSF) and whole blood (sample type is chemistry dependent).
Manufacturer(s)
Beckman Coulter Inc.