RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter Inc. recalls UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 - Product Usage:

Recalled October 7, 2020 · FDA recall Z-3028-2020 · Beckman Coulter

Hazard
The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range. · FDA Class II
Units
219 units
Sold at
Worldwide distribution - US Nationwide distribution including in the states of CA, FL and NJ. The countries of Algeria, Egypt, Philippines and Taiwan.

Description

The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range.

Product
UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 - Product Usage: System reagent for the quantitative determination of Unsaturated Iron Binding Capacity (UIBC) in human serum and plasma on Beckman Coulter AU analyzers.
Manufacturer(s)
Beckman Coulter Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov