RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter Inc. recalls iQ200 Series Urine Microscopy Analyzer- All part numbers

Recalled May 20, 2020 · FDA recall Z-1945-2020 · Beckman Coulter

Hazard
A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM. · FDA Class II
Units
3576 units
Sold at
US Distribution: Nationwide Foreign Distribution: Worldwide

Description

A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or melting of the SATA adapter cable within the computer s metal enclosure, smoke emanating from PC6, flames contained within PC6, or an inability to use the device s CD-ROM.

Product
iQ200 Series Urine Microscopy Analyzer- All part numbers
Manufacturer(s)
Beckman Coulter Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov