RecallWatchUSA

Home / Recalls / Medical Devices

Beckman Coulter Inc. recalls COULTER PrepPlus, Part Number: 286600 Product Usage: The COULTER PrepPlus, is a microprocessor-controlled pipetting

Recalled October 10, 2012 · FDA recall Z-0002-2013 · Beckman Coulter

Hazard
The recall was initiated because Beckman Coulter has confirmed that the Coulter PrepPlus may experience intermittant reagent level sense failures where the instrument may not produce a warning that a reagent vial is low or out of reagent. · FDA Class II
Units
36 units total (15 units in the US)
Sold at
Worldwide Distribution - US Nationwide and the countries of: Australia, Botswana, Canada, France, Germany, Japan, Malawi and Sweden.

Description

The recall was initiated because Beckman Coulter has confirmed that the Coulter PrepPlus may experience intermittant reagent level sense failures where the instrument may not produce a warning that a reagent vial is low or out of reagent.

Product
COULTER PrepPlus, Part Number: 286600 Product Usage: The COULTER PrepPlus, is a microprocessor-controlled pipetting and diluting system, designed for automating sample preparation or assay methods. It is capable of aspirating and dispensing liquid samples.
Manufacturer(s)
Beckman Coulter Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Beckman Coulter Inc. recalls COULTER PrepPlus, Part Number: 286600 Product Usage: The COULTER PrepPlus, is a microprocessor-controlled pipetting | RecallWatchUSA