Beckman Coulter, Inc. recalls CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU

Recalled November 5, 2025 · FDA recall Z-0412-2026 · Beckman Coulter

Hazard
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment. · FDA Class II
Units
108 units
Sold at
Worldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN, MA, ME, MI, MN, MO, NY, OR, TN, UT, WA and the countries Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Korea, Republic of, Mexico, Morocco, Netherlands, Poland, Portugal, Puerto Rico, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.

Description

Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.

Product
CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.
Manufacturer(s)
Beckman Coulter, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov