Beckman Coulter, Inc. recalls Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbishe

Recalled March 22, 2023 · FDA recall Z-1257-2023 · Beckman Coulter

Hazard
An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a remote probability of delayed results or reporting erroneous patient results. The subject nut bands were found to not meet the firm's specifications for hardness and had the incorrect material grain direction. Affected instruments were manufactured from December 28, 2017 to October 19, 2022 or had their nut bands replaced during the same period. · FDA Class II
Units
4,137 devices (990 U.S. and 3,147 OUS)
Sold at
Worldwide and Nationwide Distribution. There was government/military distribution. Foreign distribution was made to Algeria, Andorra, Australia, Azerbaijan, Bangladesh, Belarus, Belgium, Bonaire, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Curacao, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Ethiopia, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Kuwait, Kyrgyzstan, Lebanon, Libya, Lithuania, Macao, Malawi, Malaysia, Maldives, Mexico, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Swaziland, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, U.A.E, Uganda, Ukraine, United Kingdom, Uruguay, Vietnam, Yemen, Zambia, and Zimbabwe.

Description

An increased incidence of damage to the nut band from a new vendor (Boamax) has led to an increased volume of complaints from consignees regarding wash arm motion failures, system check failures, and/or assay performance leading to a remote probability of delayed results or reporting erroneous patient results. The subject nut bands were found to not meet the firm's specifications for hardness and had the incorrect material grain direction. Affected instruments were manufactured from December 28, 2017 to October 19, 2022 or had their nut bands replaced during the same period.

Product
Access Immunoassay Systems (1) Access 2 Immunoassay Analyzer (81600N); (2) Access 2 Immunoassay Analyzer, Refurbished (386220); (3) Access 2 Immunoassay Analyzer, Refurbished (A65531) ; and (4) Access 2 Section, DxC 600i Packaged(A25640).
Manufacturer(s)
Beckman Coulter, Inc.