Baxter Healthcare Corporation recalls Welch Allyn CONNEX Accessory Power Management Stand:

Recalled April 23, 2025 · FDA recall Z-1559-2025 · Baxter Healthcare

Hazard
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter. · FDA Class II
Units
114,202 units
Sold at
Worldwide - US Nationwide distribution.

Description

There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.

Product
Welch Allyn CONNEX Accessory Power Management Stand:
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov