RecallWatchUSA

Home / Recalls / Medical Devices

Baxter Healthcare Corporation recalls Voalte Patient Safety, Part Number 4.0, V4.0.000 to 4.0.401, used with the Centrella Bed

Recalled July 5, 2023 · FDA recall Z-2006-2023 · Baxter Healthcare

Hazard
Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off. · FDA Class II
Units
8 software versions
Sold at
US Nationwide distribution in the states of OR, SC, VA, TN, GA, NC, CO, NC, FL.

Description

Baxter identified a potential risk where the "safety" monitoring and "bed exit" monitoring features may enter a permanent state of alert suppression. In this state, a caregiver may not be alerted remotely when assigned patient protocols (bed exit, siderail position, brakes off, height of bed changes, head of bed angle, and/or turn reminders) are turned off.

Product
Voalte Patient Safety, Part Number 4.0, V4.0.000 to 4.0.401, used with the Centrella Bed
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Baxter Healthcare Corporation recalls Voalte Patient Safety, Part Number 4.0, V4.0.000 to 4.0.401, used with the Centrella Bed | RecallWatchUSA