RecallWatchUSA

Home / Recalls / Medical Devices

Baxter Healthcare Corporation recalls TherMax Blood Warmer Unit

Recalled November 6, 2019 · FDA recall Z-0247-2020 · Baxter Healthcare

Hazard
TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds. · FDA Class II
Units
504 units total
Sold at
Nationwide distribution to Georgia and Maryland. International distribution to Belgium, Germany, Finland, France, Greece, Italy, Luxemburg, Netherlands, Norway, Poland, Portugal, Spain, Sweden, Switzerland, United Kingdom, Australia, and New Zealand

Description

TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.

Product
TherMax Blood Warmer Unit
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov