RecallWatchUSA

Home / Recalls / Medical Devices

Baxter Healthcare Corporation recalls NaviCare Nurse Call/Voalte Nurse Call

Recalled July 6, 2022 · FDA recall Z-1306-2022 · Baxter Healthcare

Hazard
An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority. · FDA Class II
Units
283 installations
Sold at
nationwide and Canada

Description

An issue has been identified with Phillips (Emergin) and Longleaf non relay (Connexall, Vocera, Cerner) wireless integrations used with NaviCare/Voalte Nurse Call, software versions 3.9.100 through 3.9.300. Calls placed from a push button call device, such as a push button switch, call cord, or from the push buttons on a room audio station, will be canceled on the nurse call system when the call is answered at the wireless phone, regardless of the call priority.

Product
NaviCare Nurse Call/Voalte Nurse Call
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Baxter Healthcare Corporation recalls NaviCare Nurse Call/Voalte Nurse Call | RecallWatchUSA