Baxter Healthcare Corporation recalls MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541

Recalled January 31, 2018 · FDA recall Z-0411-2018 · Baxter Healthcare

Hazard
Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit. · FDA Class II
Units
186 units
Sold at
US, Germany, Sweden, Italy, Belgium, Portugal, Spain, France, Russia, Poland, Turkey, China, Thailand, Taiwan, Hong Kong, Malaysia, Korea, and Mexico.

Description

Baxter has received customer complaints regarding leakage in the albumin circuit. The leakage was caused by an inadequate adhesive connection of the tubing to the Hansen connector of the MARS Tube Set which is part of the MARS Treatment Kit.

Product
MARS Treatment Kit, Use only with MARS(R) Monitor 1TC, GAMBRO, Product Code 800541
Manufacturer(s)
Baxter Healthcare Corporation