Baxter Healthcare Corporation recalls Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expa
Recalled July 8, 2026 · FDA recall Z-2548-2026 · Baxter Healthcare
Hazard
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup. · FDA Class I
Units
10540 units
Sold at
Worldwide distribution - US Nationwide and the country of Canada.
Description
There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.
Product
Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
Manufacturer(s)
Baxter Healthcare Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →