RecallWatchUSA

Home / Recalls / Medical Devices

Baxter Healthcare Corporation recalls HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A0000

Recalled November 8, 2023 · FDA recall Z-0212-2024 · Baxter Healthcare

Hazard
Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock. · FDA Class II
Units
499 beds
Sold at
US Nationwide Distribution in the states of: CA, FL, IN, MA, NY, OH, TX, VA

Description

Progressa+ beds have the potential for static buildup from the casters and/or power drive. This static electricity could be passed to the operator through the transport handle during transport, resulting in a mild static electric shock.

Product
HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A000052, P7501A000053, P7501A000054, P7501A000055, P7501A001851, P7501A001972, P7501A001973, and P7501A001974
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Baxter Healthcare Corporation recalls HILLROM PROGRESSA+ BED, Product Codes: P7501A000016, P7501A000019, P7501A000031, P7501A000032, P7501A000033, P7501A0000 | RecallWatchUSA