Baxter Healthcare Corporation recalls Baxter Spectrum Wireless Battery Modules A/B/G/N, product code 35223, used with either of the following: 35700BAX2 (Spe

Recalled August 28, 2024 · FDA recall Z-2581-2024 · Baxter Healthcare

Hazard
The Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ infusion system is integrated with the hospital s EMR system. · FDA Class II
Units
17,377 units
Sold at
US, Nationwide

Description

The Spectrum WBM may fail to auto-document infusion status information back to the hospital's Electronic Medical Record (EMR) system. This software issue only impacts customers where the Spectrum V8 or Spectrum IQ infusion system is integrated with the hospital s EMR system.

Product
Baxter Spectrum Wireless Battery Modules A/B/G/N, product code 35223, used with either of the following: 35700BAX2 (Spectrum V8 infusion pump) or 3570009 (Spectrum IQ infusion pump) Baxter SIGMA Spectrum V8 Infusion Pump
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov