RecallWatchUSA

Home / Recalls / Medical Devices

B Braun Medical Inc recalls B. Barun Hemodialysis Bloodlines, StreamLine Bloodline Long Version, FMC. Model Number: SL-2000M2095L.

Recalled April 22, 2026 · FDA recall Z-1801-2026 · B Braun Medical

Hazard
Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure. · FDA Class I
Units
328,640 units
Sold at
Worldwide distribution - US Nationwide and the country of Canada.

Description

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Product
B. Barun Hemodialysis Bloodlines, StreamLine Bloodline Long Version, FMC. Model Number: SL-2000M2095L.
Manufacturer(s)
B Braun Medical Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov