Home / Recalls / Medical Devices
ARROW INTERNATIONAL Inc. recalls ARROW Pressure Injectable Arrowg+ard Blue Advance Two-Lumen Midline Kit, 5.5 Fr, 15 cm, .018 Inch dia., REF ASK-41552-JH
Recalled October 26, 2022 · FDA recall Z-0102-2023 · ARROW INTERNATIONAL
Hazard
Mislabeling: the corrugate label, lidstock, banner card, and peel sticker accompanying the affected product incorrectly displays a double lumen 5.5 French catheter (and related priming volumes and kit contents for product code ASK-41552-JHVA1), instead of the packaged single-lumen 4.5 French catheter (and related priming volumes and kit contents for product code ASK-41541-JHVA1). · FDA Class II
Units
18 devices
Sold at
US Nationwide distribution in the state of Florida.
Description
Mislabeling: the corrugate label, lidstock, banner card, and peel sticker accompanying the affected product incorrectly displays a double lumen 5.5 French catheter (and related priming volumes and kit contents for product code ASK-41552-JHVA1), instead of the packaged single-lumen 4.5 French catheter (and related priming volumes and kit contents for product code ASK-41541-JHVA1).
Product
ARROW Pressure Injectable Arrowg+ard Blue Advance Two-Lumen Midline Kit, 5.5 Fr, 15 cm, .018 Inch dia., REF ASK-41552-JHVA1. For peripheral access to the venous system for intravenous therapy, blood sampling, infusion, and pressure injection of contrast media.
Manufacturer(s)
ARROW INTERNATIONAL Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →