RecallWatchUSA

Home / Recalls / Medical Devices

Angiodynamics, Inc. recalls VenaCure EVLT KIT WITH SPOTLIGHT OPS SHEATH 55CM Ref./ Catalog No.: EVLT55OPS (each box contains 5 Procedure Kits). F

Recalled October 5, 2022 · FDA recall Z-1819-2022 · Angiodynamics

Hazard
OPS Fiber packaged within specific VenaCure EVLT Procedure Kits does not meet the required dimensional specification, may result in damage to the sheath, including separation or the inability to move the fiber within the sheath. A damaged sheath will necessitate removal from the patient, which may result in procedural delay and/or surgical intervention · FDA Class II
Units
40 units
Sold at
US Nationwide distribution in the states of FL, NC, NJ, NM, NY, PA, TX, VA.

Description

OPS Fiber packaged within specific VenaCure EVLT Procedure Kits does not meet the required dimensional specification, may result in damage to the sheath, including separation or the inability to move the fiber within the sheath. A damaged sheath will necessitate removal from the patient, which may result in procedural delay and/or surgical intervention

Product
VenaCure EVLT KIT WITH SPOTLIGHT OPS SHEATH 55CM Ref./ Catalog No.: EVLT55OPS (each box contains 5 Procedure Kits). For use in the treatment of varicose veins.
Manufacturer(s)
Angiodynamics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Angiodynamics, Inc. recalls VenaCure EVLT KIT WITH SPOTLIGHT OPS SHEATH 55CM Ref./ Catalog No.: EVLT55OPS (each box contains 5 Procedure Kits). F | RecallWatchUSA