Home / Recalls / Medical Devices
American Contract Systems, Inc. recalls Vasc. Open CVOR SJH, REF SJVO56B
Recalled November 1, 2023 · FDA recall Z-0177-2024 · American Contract Systems
Hazard
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals. · FDA Class II
Units
84 cases
Sold at
US Nationwide distribution in the states of FL, IA.
Description
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
Product
Vasc. Open CVOR SJH, REF SJVO56B
Manufacturer(s)
American Contract Systems, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →