RecallWatchUSA

Home / Recalls / Medical Devices

American Contract Systems Inc. recalls Medical convenience kits Item Number/Description EVLH67F LEFT HEART CATH PACK FDCL22R CATH LAB ANGIO IHCC03X CATH

Recalled October 22, 2025 · FDA recall Z-0175-2026 · American Contract Systems

Hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. · FDA Class II
Units
32,433 total
Sold at
US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Description

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Product
Medical convenience kits Item Number/Description EVLH67F LEFT HEART CATH PACK FDCL22R CATH LAB ANGIO IHCC03X CATH CARDIAC PACK TNCC02AD CARDIAC CATH TRINITY UICC27AG CARDIAC CATH ANGIO PACK (PS 923077) UIRD89AH RADIAL PACK (PS 56573)
Manufacturer(s)
American Contract Systems Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov