Home / Recalls / Medical Devices
American Contract Systems Inc. recalls Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PAC
Recalled October 22, 2025 · FDA recall Z-0171-2026 · American Contract Systems
Hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. · FDA Class II
Units
32,433 total
Sold at
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Description
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Product
Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25AJ LAPROSCOPY PACK - 205958, FYLA38E LAPCHOLAPPY PACK, IHDA88AH DIVINCI GYN URO, IHLC11AE LAP GENERAL PACK, MHLP45AJ LAPAROSCOPY PACK - 206005, MHRB88AO ROBOTIC PACK - 242561, RCLC51C LAP CHOLE RCH 0246969, SMLC53C LAP CHOLE PACK, UDLC17AE DSC LAP CHOLE PACK - 206062, UIRU84AE ROBOTIC UROLOGY PACK (PS 060571)
Manufacturer(s)
American Contract Systems Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →