RecallWatchUSA

Home / Recalls / Medical Devices

Alcon Research, Ltd. recalls Alcon Single-Use Curved Jewelers/Iris Bipolar Forceps REF 8065129101 Sterile RX only; Alcon Single-Use Straight/Iris Bip

Recalled December 17, 2014 · FDA recall Z-0527-2015 · Alcon Research

Hazard
The product is sold and labeled as EO sterilized, however the units were not sterilized prior to shipment to customers. · FDA Class II
Units
2321
Sold at
Worldwide distribution: US (nationwide) and countries of: Indonesia, South Africa, Switzerland, Brazil, Kazakhstan, and Pakistan.

Description

The product is sold and labeled as EO sterilized, however the units were not sterilized prior to shipment to customers.

Product
Alcon Single-Use Curved Jewelers/Iris Bipolar Forceps REF 8065129101 Sterile RX only; Alcon Single-Use Straight/Iris Bipolar Forceps REF 8065129501 Sterile RX only. Usage: Electrosurgical instrument for ophthalmic surgery.
Manufacturer(s)
Alcon Research, Ltd.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov