RecallWatchUSA

Home / Recalls / Medical Devices

Alcon Research LLC recalls Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Mod

Recalled November 5, 2025 · FDA recall Z-0304-2026 · Alcon Research

Hazard
Due to incomplete seals in the pouch which provide the sterile barrier. · FDA Class II
Units
577
Sold at
Worldwide distribution: US (nationwide): AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC,SD,TN, TX, VA, WA, WI, and WV; and OUS (foreign) to countries of: Argentina, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belgium, Brazil, Brunei, Canada, Chile, China, Colombia, Croatia, Czechia, Denmark, Finland, France, Germany, Greece, Guadeloupe, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Korea, Kosovo, Luxembourg, Malaysia, Mexico, Moldova, Montenegro, Morocco, Netherlands, New Zealand, Pakistan, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Russian Federation, Saudi Arabia, Serbia, Slovakia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom,

Description

Due to incomplete seals in the pouch which provide the sterile barrier.

Product
Brand Name: Alcon Laboratories, Inc. Product Name: Accessory devices for Unity and Constellation surgical consoles Model/Catalog Number: Many - See 'Attachment 1 - Affected Product List' 8065000250 UNITY 27+DS HYPERVIT 30K VIT PROBE 8065753168 UNITY 25+ 7.5K ANTERIOR VIT PROBE 8065753179 UNITY 25+ HYPERVIT 30K VIT PROBE Software Version: N/A Product Description: These products are standalone accessory devices meant for vitrectomy surgical procedures. They interface with a corresponding Alcon surgical console. Component: N/A
Manufacturer(s)
Alcon Research LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov