November 19, 2025Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls MAGNETOM Cima.X Upgrade. Model Number: 11689304.
Hazard: There is a potential for an ice blockage to form or currently exist within the primary and/or the auxiliary venting paths of the magnet helium containment venting system. In the event of a quench when both venting paths are blocked, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
November 12, 2025Medical DevicesFDA
Beckman Coulter Inc. recalls SYNCHRON Systems Direct LDL Cholesterol (LDLD) Reagent, REF: 969706
Hazard: Beckman Coulter has identified that there is an increased probability Synchron Systems Direct LDL Cholesterol (LDLD REF 969706) reagent lot M404166 may generate false high results. The issue was discovered internally and confirmed on August 19, 2025 by Beckman Coulter.
November 12, 2025Medical DevicesFDA
B-K Medical A/S recalls bk5000 Ultrasound System w/battery; Model No. 2300-61; System, Imaging, Pulsed Doppler, Ultrasonic
Hazard: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
November 12, 2025Medical DevicesFDA
B-K Medical A/S recalls bk3500 Ultrasound System w/battery; Model No. 2300-41; System, Imaging, Pulsed Doppler, Ultrasonic
Hazard: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
November 12, 2025Medical DevicesFDA
B-K Medical A/S recalls bk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, Ultrasonic
Hazard: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
November 12, 2025Medical DevicesFDA
B-K Medical A/S recalls UUltrasound System 2300, bkActiv w/battery; Model No. 2300-11; System, Imaging, Pulsed Doppler, Ultrasonic
Hazard: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
November 12, 2025Medical DevicesFDA
B-K Medical A/S recalls Ultrasound System 1300, bkSpecto w/battery; Model No. 1300-21 or 1300-S1; System, Imaging, Pulsed Doppler, Ultrasonic
Hazard: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
November 12, 2025Medical DevicesFDA
B-K Medical A/S recalls 1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focu
Hazard: GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
November 12, 2025Medical DevicesFDA
Elekta, Inc. recalls Brand Name: Leksell Stereotactic System Product Name: Leksell Vantage Stereotactic System Software Version: n/a Produ
Hazard: Due to increased risk that paint flakes from the engravings may separate from the device and there is the possibility that the flake may inadvertently enter the wound
November 12, 2025Medical DevicesFDA
Total Joint Orthopedics, Inc. recalls Brand Name: Klassic BiPolar Head Model/Catalog Number: 3205.41.022 Klassic BiPolar Head, 41 mm OD, 22 mm Head; 320
Hazard: Potential packaging issue that could compromise the sterile barrier. Defect may result in loss of sterility of the device, presenting potential risk of infection if device is implanted.
November 12, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter UniCel DxI 800, Part Numbers: 973100, A71456; Photometric Chemistry Analyzer
Hazard: An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
November 12, 2025Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Hazard: An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.
November 12, 2025Medical DevicesFDA
Tandem Diabetes Care, Inc. recalls Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both
Hazard: Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
November 12, 2025Medical DevicesFDA
Mozarc Medical US LLC recalls MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Safety Components: 12 Fr/Ch (4.0 mm
Hazard: A potential issue with the seal integrity of header bag packaging.
November 12, 2025Medical DevicesFDA
Mozarc Medical US LLC recalls MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 24
Hazard: A potential issue with the seal integrity of header bag packaging.
November 12, 2025Medical DevicesFDA
Mozarc Medical US LLC recalls MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Straight Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20
Hazard: A potential issue with the seal integrity of header bag packaging.
November 12, 2025Medical DevicesFDA
Mozarc Medical US LLC recalls MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 20 cm
Hazard: A potential issue with the seal integrity of header bag packaging.
November 12, 2025Medical DevicesFDA
Mozarc Medical US LLC recalls MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 16 cm
Hazard: A potential issue with the seal integrity of header bag packaging.
November 12, 2025Medical DevicesFDA
Mozarc Medical US LLC recalls MAHURKAR" Acute High Pressure Triple Lumen Catheter Tray Curved Extensions with IC Components: 12 Fr/Ch (4.0 mm) x 13 cm
Hazard: A potential issue with the seal integrity of header bag packaging.
November 12, 2025Medical DevicesFDA
Integra LifeSciences Corp. (NeuroSciences) recalls TruDi¿ Navigation System - Multi Instrument Adapter; - Manufacturer's Product Number (Catalog Number): EM-2025-000F Rev
Hazard: It was determined that when the navigation system is configured with specific revision combinations of the Multi Instrument Adapter together with the Patient Tracker the system may not meet its specified accuracy for visual verification of device location within the patient s anatomy.
November 12, 2025Medical DevicesFDA
Zimmer, Inc. recalls Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 46 mm O.D., Size G, Bearing, Model/Ca
Hazard: Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
November 12, 2025Medical DevicesFDA
Zimmer, Inc. recalls Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Ca
Hazard: Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E¿ Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
November 12, 2025Medical DevicesFDA
Prismatik Dentalcraft, Inc. recalls Brand Name: Inclusive Product Name: Inclusive Titanium Screw compatible with: Biomet 3i" Certain 3.4/4.1/5.0/6.0 mm Mo
Hazard: Incorrect titanium screw, packaged with dental implant.
November 12, 2025Medical DevicesFDA
Spacelabs Healthcare, Ltd. recalls Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Des
Hazard: Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record
November 12, 2025Medical DevicesFDA
Integra LifeSciences Corp. (NeuroSciences) recalls CODMAN Disposable Perforator 9mm. Cranial Perforator.
Hazard: Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.
November 12, 2025Medical DevicesFDA
Integra LifeSciences Corp. (NeuroSciences) recalls CODMAN Disposable Perforator 11mm. Cranial Perforator.
Hazard: Expansion of recall of size 14 mm Codman Disposable Perforators for inadequate welds to size 11 mm and 9 mm Codman Disposable Perforators out of an abundance of caution.
November 12, 2025Medical DevicesFDA
OSSTEM Implant Co., Ltd. recalls Osstem Dental X-Ray System T2Plus. Product Model Numbers T2-C-P and T2-CS-P.
Hazard: Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted: Certification Statement: This electronic product complies with all applicable performance standards prescribed under 21 CFR Subchapter J, as enforced by the U.S. Food and Drug Administration at the date of manufacturing.
November 12, 2025Medical DevicesFDA
OSSTEM Implant Co., Ltd. recalls Osstem Dental X-Ray System T2. Product Model Numbers T2-C and T2-CS.
Hazard: Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted: Certification Statement: This electronic product complies with all applicable performance standards prescribed under 21 CFR Subchapter J, as enforced by the U.S. Food and Drug Administration at the date of manufacturing.
November 12, 2025Medical DevicesFDA
OSSTEM Implant Co., Ltd. recalls Osstem Dental X-Ray System T1. Product Model Numbers T1-C and T1-CS.
Hazard: Osstem Implant Co. Ltd, is voluntarily initiating a medical device correction involving the current labeling for this electronic product may be missing a required Certification Statement under 21 CFR Subchapter J (Radiological Health Regulations). Specifically, the following statement may have been omitted: Certification Statement: This electronic product complies with all applicable performance standards prescribed under 21 CFR Subchapter J, as enforced by the U.S. Food and Drug Administration at the date of manufacturing.
November 5, 2025Medical DevicesFDA
Microgenics Corporation recalls Thermo scientific MAS Omni CARDIO, Liquid Assayed Integrated Cardiac Control, Model/Catalog Number: OCRD-L, OCRD-101, OC
Hazard: The firm received complaints from customers reporting vial-to-vial variability when using the control with high sensitivity Troponin I assays. This defect may lead to as delay in patient results for hs Troponin I assays.