March 6, 2024Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE BRONCHOSCOPY TRAY, REF DYNDL1350A
Hazard: Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
March 6, 2024Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE kits and trays, labeled as: a) SUCTION CATH 10FR KIT W/SALINE, REF DYND41470SA, DYND41470SAH; b) TRACH CARE
Hazard: Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
March 6, 2024Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE TRAY care and cleaning kits and trays for Tracheostomy and Stomas, labeled as: a) STOMA & TRACH CLEANING KIT,
Hazard: Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
March 6, 2024Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls MEDLINE TRAY TRACH BASIC SOLUTION, REF DYND40511
Hazard: Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.
March 6, 2024Medical DevicesFDA
Ventec Life Systems, Inc. recalls VOCSN Patient Breathing Package (Pediatric, Active, Oxygen, Blue) REF PRT-01013-000 with pouch REF PRT-00802-001). Used
Hazard: The bonded spiral wrap may detach before or during ventilation due to a manufacturing issue. Detachment of the spiral wrap can compromise structural integrity, functionality and/ or performance resulting in occlusions or leaks of the breathing circuit which may cause serious injury or death. The risk of injury or death is increased for ventilator dependent patients.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8004
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Variable Volume Oxygen Reservoir Tubing, Adult Mask, REF 2K8017
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0 m) Oxygen Reservoir Tubing, Adult Mask, CO2 Detector, REF 2K8005C2
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, Expiratory Filter, REF 2K8005F
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, Adult Mask, REF 2K8005
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Adult Mask, CO2 Detector, REF 2K8004C2
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, Expiratory Filter, REF 2K8004F
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, 40" (1.0m) Oxygen Reservoir Tubing, without Mask, REF 2K8001
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
March 6, 2024Medical DevicesFDA
Vyaire Medical recalls AirLife Adult Manual Resuscitator, Oxygen Reservoir Bag, without Mask, REF 2K8000,
Hazard: Risk of hypoventilation caused by duckbill and ring broken/disassembled, due to faulty mold/tooling producing out of specification components. Device failure may cause improper ventilation or no ventilation that may result in hypoventilation or hypoxia, and potentially lead to death. The mold/tooling used to produce the affected component was replaced in 2017.
February 28, 2024Medical DevicesFDA
PTW-FREIBURG recalls Software VERIQA (S070031), version 2.0 and 2.1- Software package for display, evaluation and digital processing of medic
Hazard: Software: If the user excludes voxels from the Gamma calculation that are below a dose threshold (Suppress gamma calculation), the Gamma Passing Rate (GPR) calculated for individual ROIs is not correct. The GPR calculated by VERIQA overestimates the correct GPR. The evaluation can therefore show false positive results.
February 28, 2024Medical DevicesFDA
MEDLINE INDUSTRIES, LP - Northfield recalls These wipes are saturated with 0.65% bleach solution. These wipes are antimicrobial, disinfectant with sporicidal, viruc
Hazard: There is stability concern for products dated with a 24-month shelf life. Product is safe and effective up to 12 months shelf life.
February 28, 2024Medical DevicesFDA
W.L. Gore & Associates, Inc. recalls GORE VIABIL Short Wire Biliary Endoprosthesis, Product labeled as (1) Catalog Number VSWVH1008, 10 mm x 8 cm; with ho
Hazard: Devices without transmural drainage holes may be mislabeled as devices with transmural drainage holes, or vice versa. Potential health impact includes extending procedure time, unplanned device removal, secondary procedure, cholangitis, cholecystitis, and/or pain. The GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis is a flexible, self-expanding endoprosthesis designed to maintain luminal patency of the common bile duct which may be obstructed. The endoprosthesis is radially compressed and secured onto the distal end of a catheter-based delivery system for deployment. The catheter provides a means for accessing and implanting the GORE¿ VIABIL¿ Short Wire Biliary Endoprosthesis at the target site in the biliary tract.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 7 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surg
Hazard: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 5 with a Certeray generator-To perform image guidance in diagnostic, interventional, and minimally invasive surg
Hazard: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion 3 with a Certeray generator -To perform image guidance in diagnostic, interventional, and minimally invasive sur
Hazard: Generator may fail due to a potential short circuit in the Printed Circuit Board Assembly (PCBA) in PoInt EVR (Power Invertor). If a short circuit occurs, it will cause the fuses to trip and the system will become non-functional due to power loss, resulting in a potential delay of procedure or termination of procedure.
February 28, 2024Medical DevicesFDA
Abiomed, Inc. recalls Automated Impella Controller (AIC), Product No. 0042-0000-US, 0042-0010-US, 0042-0040-US. Used with Impella Ventricular
Hazard: Retrospective reporting for the release of Technical Bulletin IMP-2643 AIC Version 8.5 Software Update Available. The update resolved an issue in version 8.4 where a pump was not recognized by the AIC.
February 28, 2024Medical DevicesFDA
MRP, LLC dba Aquabiliti recalls AQUABILITI AQUASTAT 0.9% Sodium Chloride Injection syringe, USP, 5mL in 12mL, REF 2T0805; and 3mL in 12mL, REF 2T0807, i
Hazard: The saline flush syringes may fail to meet USP criteria for stated sodium chloride content after 12 months.
February 28, 2024Medical DevicesFDA
IMPULSE DYNAMICS (USA) INC recalls OPTIMIZER model CCM X11 implantable pulse generator (IPG) devices - Smart Mini and Lite
Hazard: OPTIMIZER devices may cease to deliver CCM therapy if the device incorrectly detects a charging error. This may cause patients to experience heart failure symptoms similar to before implantation of the device.
February 28, 2024Medical DevicesFDA
FUJIFILM Healthcare Americas Corporation recalls Synapse CV 6. with AR. A web-based application as the primary user interface for the processing of medical images.
Hazard: The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
February 28, 2024Medical DevicesFDA
FUJIFILM Healthcare Americas Corporation recalls Synapse Cardiology PACS V7.3.0, V7.2, V7.1, V7.0. A web-based application as the primary user interface for the proces
Hazard: The LV Mass (2D Bullet) equation may be calculated incorrectly, resulting in a variance in the mass of the left ventricle (LV Mass). If used as a main factor for diagnosis, there is a risk of misdiagnosis or incorrect treatment plan of a patient, resulting in long-term health consequences or serious deterioration of health.
February 28, 2024Medical DevicesFDA
3M Company - Health Care Business recalls 3M Specialty Blade Assembly, REF 9660,70-2011-8491-1, 7100213026, used with used with 3M Surgical Clipper Catalog Number
Hazard: 3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41¿C (105.8¿F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60¿C (140¿F) and the Model 9660 blade maximum temperature may reach 77¿C (170.6¿F)."
February 28, 2024Medical DevicesFDA
3M Company - Health Care Business recalls 3M Specialty Blade Assembly, REF 9660,70-2011-8462-2, 7100213114, used with used with 3M Surgical Clipper Catalog Number
Hazard: 3M has added the following cautionary statement regarding the use of the product: "Caution - To reduce the risk of burns, which if not avoided may result in minor injury: " Do not use on injured skin. " Do not leave clipper running unused in air without applying to skin for more than 1 minute as blade temperature may exceed 41¿C (105.8¿F). " Do not leave the clipper blade applied to the same location on the patient s skin for longer than 1 second as potential injury may occur; continuous motion of the clipper during use is required. Note: If clipper is left running unused in air for 10 minutes without applying to skin, the Model 9690 blade maximum temperature may reach 60¿C (140¿F) and the Model 9660 blade maximum temperature may reach 77¿C (170.6¿F)."
February 28, 2024Medical DevicesFDA
W&H DENTALWERK BUERMOOS GMBH recalls W&H Surgical handpiece S-12. For Surgical treatment of dental hard tissue Model Number: 30061000
Hazard: 1:2 speed increasing surgical handpieces have been incorrectly laser marked during production. The specific laser marking should read 1:2, but instead it reads 1:1 on the body of the handpiece
February 28, 2024Medical DevicesFDA
Olympus Corporation of the Americas recalls ESG PK CUTTING FORCEPS, 5MM, 33CM
Hazard: The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.
February 28, 2024Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the DIMMs (Dual In-line Memory Modules) used
Hazard: Dual In-Line Memory Modules (DIMMs) may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.