Home / Brands / Spacelabs Healthcare
Spacelabs Healthcare Recalls
47 recalls on record · June 27, 2012 to August 19, 2026 · Medical Devices
Recalls by year
12
13
14
15
16
17
18
19
20
22
23
24
25
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
March 4, 2015Medical DevicesFDA
Spacelabs Healthcare Inc recalls Spacelabs Healthcare XPREZZON Bedside Monitor, Model 91393. The Spacelabs Healthcare XPREZZON Bedside Monitor passively
Hazard: Spacelabs Healthcare XPREZZON Bedside Monitors, Model 91393, are recalled because the firm has received multiple reports of XPREZZON monitors failing to boot up or returning to factory default configuration settings following power on or reset.
October 1, 2014Medical DevicesFDA
Spacelabs Healthcare Inc recalls Spacelabs Medical Patient Monitor is a Physiological, Patient monitor With Arrhythmia Detection or Alarms. The Models ar
Hazard: The clinician did not realize the monitor was in END CASE mode and was expecting audible alarms. Visual alarms continue to work normally. The firm performed a retrospective analysis of all recalls since 2011 and discovered 222 additional serial numbers were missed and were not included in the previous recall under RES 59456 (recall number Z-3032-2011).
September 10, 2014Medical DevicesFDA
Spacelabs Healthcare Inc recalls ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Physiological patient monitor. Used in a hospital env
Hazard: The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions. This is an expansion of recall Z-0871-2014 because additional serial numbers were identified as being part of this recall action as part of the firm's retrospective analysis of all recalls since 2011.
July 23, 2014Medical DevicesFDA
Spacelabs Healthcare Inc recalls Spacelabs Healthcare Ultraview SL Multigas Module, Model 92518 (Software V1.00.00-09). Provides a means of monitoring a
Hazard: With this software version, the Minimum Alveolar Concentration (MAC) value displayed is extremely high (20.1) and physiologically invalid.
July 2, 2014Medical DevicesFDA
Spacelabs Healthcare Inc recalls Spacelabs Healthcare Qube Compact Monitor, Model 91390. Intended to alert the user to alarm conditions that are repor
Hazard: Spacelabs Healthcare is recalling Qube Compact Monitor, Model 91390, due to a potential chance of losing network connection after switching from hardwired Ethernet connection to using the Qube Docking Station without powering down the device. The bedside monitor and all of its alarms will continue to function normally. No one has been injured as a result of this issue.
June 25, 2014Medical DevicesFDA
Spacelabs Healthcare Inc recalls Spacelabs Healthcare Ventilator Flexport Interface, Model 90436A-07
Hazard: Spacelabs Healthcare is voluntarily recalling the Hamilton Galileo Ventilator Flexport, Model 90436A-07, where the monitored Minute Volumes (Vmin) has been reported at one time to reach ten times the actual value on the bedside monitor.
June 11, 2014Medical DevicesFDA
Spacelabs Healthcare Inc recalls Spacelabs Medical Ultraview SL Command Module, Model 91496. Arrhythmia Detector and Alarm. The Multi-parameter Modu
Hazard: A software anomaly causes temperature alarm limits to change to limits which are not physiologically possible and result in nuisance false positive alarms.
April 2, 2014Medical DevicesFDA
Spacelabs Healthcare, Llc recalls Spacelabs Healthcare G2 Clinical Access Mislabeled CDs for PN: 063-1829-11 Rev A and Update Kit, PN: 040-1565-10. Use
Hazard: Spacelabs has voluntarily recalled G2 Clinical Access Software CD because the programs installed on the Clinical Access CD are incorrect and it will prevent the customer from installing Clinical Access.
February 5, 2014Medical DevicesFDA
Spacelabs Healthcare, Llc recalls ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with either adult or
Hazard: The AriaTele Model 96281 with SpO2 monitoring Option C is recalled because the AriaTele (transmitter) display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions.
November 27, 2013Medical DevicesFDA
Spacelabs Healthcare, Llc recalls Spacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmia detection or al
Hazard: Spacelabs Healthcare qube Compact Patient Monitor, Model 91390, may fail to power ON due to incorrect programmed microcontroller component.
November 20, 2013Medical DevicesFDA
Spacelabs Healthcare, Llc recalls Spacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with arrhythmia de
Hazard: Spacelabs Healthcare elance Vital Signs Monitor with Option S, model 93330, is recalled because the monitor will not permit connection of any of the Spacelabs TruLink line of SpO2 sensors.
January 23, 2013Medical DevicesFDA
Spacelabs Healthcare, Llc recalls Spacelabs Smart Disclosure System, Model 92810, is an integrated component of the Intesys Clinical Suite (ICS) G2, Versi
Hazard: Displaying or printing of an incorrect waveform for the time indicated in the printout or display in the Spacelabs Healthcare Intesys Clinical Suite (ICS) G2, Version 4.03, Clinical Access portion of Smart Disclosure, Model 92810.
November 28, 2012Medical DevicesFDA
Spacelabs Healthcare, Llc recalls qube Compact Monitor (Patient Monitor) 91390. The monitor is 26.2 cm x 31.5 cm x 13.2 cm and weighs 4.1 kg (10.3 inches
Hazard: Spacelabs Healthcare has learned through several reports that the input circuit may fail and the monitor will go on battery mode. The monitor will function normally until the batteries become depleted and then the monitor will shut down. In addition the monitor may fail to turn on due to an internal short.
November 14, 2012Medical DevicesFDA
Spacelabs Healthcare, Llc recalls The touchscreen display (model 94267) designed for use with the XPREZZON bedside monitor (model 91393) is width-452 mm (
Hazard: Potential failure of Spacelabs Healthcare Touchscreen Display, Model 94267 where the alarm lights failed to illuminate upon installation.
September 26, 2012Medical DevicesFDA
Spacelabs Healthcare, Llc recalls Ultraview SL 91493 Integrated Module Housing. PCIS patient monitors is intended to sound various alarm tones that are
Hazard: There is a potential failure of the Integrated Module Housing causing the patient module parameters to drop off the monitor display or the patient module will fail to sign on when inserted into the housing.
July 25, 2012Medical DevicesFDA
Spacelabs Healthcare, Llc recalls Spacelabs Medical Xprezzon Bedside Monitor, Model 91393, installations with Option C. It is a patient monitor with arrhy
Hazard: There were reports that a certain combination of key presses will cause the module to disconnect from the Xprezzon bedside monitor, model 91393, and the patient parameters will drop from the display. The parameters return after power cycling the monitor.
June 27, 2012Medical DevicesFDA
Spacelabs Healthcare, Llc recalls The product is Pathfinder SL, version 1.6.0.3057 software used with the Spacelabs Healthcare Pathfinder SL Holter Analyz
Hazard: The firm received reports of the Pathfinder LS software crashing after editing a batch of Holter Analyzer records. The software will stop with an error message requiring a restart. In addition, QTc values will not update after moving the marker.