October 30, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc. recalls Acuson SC2000 volume imaging ultrasound systems at software version 3.0. Mfg name: Siemens Medical Solutions USA, In
Hazard: Software issue results in inaccurate 3D image may result in ablation in wrong location.
September 4, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Axiom Artis systems. C-arm angiographic x-ray system x-ray, angiographic system.
Hazard: The firm became aware of a potential issue with Axiom Artis systems installed by a third party installer that may involve incorrect channel nuts being used.
September 4, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls PRIMUS, ONCOR, ARTISTE Linac systems with component Tabletop, 550TxT, TT-A s/n 1001-1233 and Tabletop 550TxT, TT-S s/n 1
Hazard: Siemens has become aware of a customer complaint concerning the rail threaded holes of the 550TxT treatment tabletop. Due to a manufacturing error, the screws holding the side rails in position on the tabletop might break off.
August 21, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens syngo.plaza SW version VA20 with Mammography license. image processing, radiological system
Hazard: Siemens issued a Customer Safety Notice to inform customers of a potential problem when using syngo.plaza. In some instances, it can happen that the laterality information in mammography images is not displayed and these images may not be properly aligned to mammography segments.
August 21, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens AXIOM Artis zeego x-ray, angiographic system
Hazard: Siemens issued a customer safety notice about the overlay of system information after a functional check of the brakes on the Artis Zeego system.
August 14, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Mobiletta Mira Mobile X-ray System
Hazard: Siemens issued a Field Safety Alert about the potential risk to users when operating the Mobilett Mira. When a Mobilett Mira is regularly exposed to extensive mechanical strain (driven over high thresholds for doors or elevators), it is possible that one of its main wheel shafts could break because of material fatigue, resulting in the chassis falling on the ground on the affected wheel side. This
August 7, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls syngo Dynamics Information System (sDIS) versions 9.5 and VA10A. The syngo¿ Dynamics system is intended for the accep
Hazard: Siemens initiate this recall due to a potential issue when using the sDIS matching rules of the syngo Dynamics Information System version 9.5 or VA10A with modality device configured to end study at association close. Images or DICOM SR objects sent from a modality device to syngo Dynamics may not be saved when the modality is configured at syngo Dynamics to end study at association close, and whe
August 7, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens Luminos dRF Fluoroscopic X-Ray system Fluoroscopic X-Ray system
Hazard: Siemens became aware of an unintended behavior when using the Luminos dRF with software version VD10A. The possibility exists that a software error may occur causing the table to tilt into an upright position and may collide with the floor. Four consignees were notified of this recall by Update Instruction XP067/12/S -- all four units have been corrected.
July 31, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Axiom Aristos MX Solid State Flat Panel Digital Imager, Model number 5895003. The device is used for diagnostic X-ray im
Hazard: Siemens issued a Field Safety Notice about the potential hazard to patients or operators when using the function "table top float" of the AXIOM Aristos MX. When repositioning a patient lying on the table top and releasing the table top, it is possible that the fingers of the patient or operator are between the table top and the detector tray and if the table top is moved, the fingers can be squee
July 31, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls ARTIS zee and zeego x-ray, angiographic system with Software Artis VC20x/VC21A that have Patch 7 and a 3D and DSA licens
Hazard: Siemens issued a customer safety advisory notice and field safety corrective action for a potential issue on Artis systems with Artis VC20x/VC21A that have Patch 7 and a 3D and DSA license installed. The issue affects-DSA rotational acquisition. The restriction is related to any rotational 3D-acquisition program with real-time subtraction (e.g. 3D DSA). During the fill phase of the rotational aqui
July 24, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens AXIOM Artis zee/zeego systems equipped with generator A100 Plus Product Usage: Angiographic x-ray system
Hazard: During the course of product monitoring, Siemens became aware of a tolerance problem in the power supply of the generator control above a specific value, thus, the generator may fail.
July 17, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens Mammomat Inspiration - full-field digital mammography Product Usage: Mammography
Hazard: Firm became aware of an unintended behavior when using the Mammomat Inspiration. The possibility exists that a malfunction may occur during Stereo or Tomo acquisitions if the tomo clutch was incorrectly adjusted. Although the possibility is rare, the system could move while a patient's breast is compressed resulting in a possible bruise. Firm released Update Instructions XP031/12/S for this iss
July 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens MEVATRON series, PRIMART (limited release)PRIMUS and/or ONCOR or ARTIST Linac Systems with BEAMVIEW capability.
Hazard: The information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in mistreatment of patients.
June 26, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls ARTISTE with syngo RT Therapist version 4.3 Product Usage: The intended use of the SIEMENS branded ARTISTE family o
Hazard: The potential for a safety issue, for the image quality degradation, that is addressed by the Customer Safety Advisory Notice when verifying the patient position based on 2D images that were acquired as single-exposure MV or In-Line kView images using the syngo RT Therapist 4.3.
May 8, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Software correction applies to customers who use the RT Therapist RTT4.2.108 and RT Therapist Connect RTT 4.2.108 on
Hazard: Software update to fix multiple safety related issues.
April 24, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens syngo WorkFlow SLR, Radiological Information System, Version VA31A SP1/SP2/SP3 The information system syngo(
Hazard: Under certain conditions, when using the Search functionality, the potential exists for the wrong patient's Interactive Documents to display when viewing a signed report in the single patient view. This can occur when the Interactive Document is viewed or auto displayed, dismissed by the user, and then a different patient and signed exam is selected in single patient view. The Notes Viewer will
April 17, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Sele
Hazard: Siemens has become aware of a potential malfunction and possible hazard to patients and staff when using the Iconos R200 C20, Iconos R200 T20, Iconos R100, Iconos MD, Iconos RHD, Luminos RF Classic, Luminos Fusion or Luminos Select. This malfunction is possible in cases where the table vertical lift movement is obstructed by a solid object which could potentially result in collision. This can re
April 17, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc. recalls Siemens brand ACUSON SC2000 Ultrasound System, Model 10433816, hardware that contains version 4 of the MPI board 1043420
Hazard: When using the V5M transducer, the display of the temperature on the ACUSON SC2000 system can be lower than the actual temperature of the V5M transducer. This occurs only if the system hardware contains revision 4 of the MPI Board. There is a potential to cause esophageal burns in patients.
April 17, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls ELEKTA MOSAIQ SW version below 2.41 Product Usage: The intended use of the SIEMENS branded ARTISTE and ONCOR famil
Hazard: A potential safety risk with the MOSAIQ Oncology Information System that has been identified by the manufacturer IMPAC Medical Systems, Inc.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 10, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE
Hazard: Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
April 3, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system
Hazard: The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.
April 3, 2013Medical DevicesFDA
Siemens Medical Solutions USA, Inc recalls Siemens Axiom Artis zeego systems with software version VC14, VC20 and VC21 in conjunction with specific technical confi
Hazard: In the course of the firm's product monitoring activities, a potential risk for patient or operators was determined during the operation of an Artis zeego system with software revision VC14, VC20 and VC21 in conjunction with a specific technical configuration which cannot be completely executed.