Home / Brands / Philips North America
Philips North America Recalls
530 recalls on record · April 5, 2017 to July 1, 2026 · Medical Devices
Recalls by year
17
18
19
20
21
22
23
24
25
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882476 - Gemini TF Big Bore PET/CT
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882446 - Vereos PET/CT
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 882442/882456 - Ingenuity TF PET/CT
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728333 - Spectral CT:
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor 728332 - IQon Spectral CT : IQon Spect
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Ingenuity Core, Ingenuity Core 128, Ingenui
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT s
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT s
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT s
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and MX16 Mo
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
July 7, 2021Medical DevicesFDA
Philips North America Llc recalls Continuous CT Fluoroscopy DVI Ceiling Mount CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT s
Hazard: Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
June 16, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Ambition S - Model Number: 781359 Diagnostic MR system.
Hazard: The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
June 16, 2021Medical DevicesFDA
Philips North America Llc recalls Ingenia Ambition X - Model Number: 781356, Diagnostic MR system. ***Updated November 2023*** Added Model Number 782109
Hazard: The MR magnet is equipped with 4 lifting blocks, which are positioned on top of the magnet. These blocks are used to allow the magnets to be moved in or out by a lifting crane, e.g. during decommissioning or moving activities. A mechanical issue was detected on these hoisting interface blocks on top of the magnet.
June 9, 2021Medical DevicesFDA
Philips North America Llc recalls BRILLIANCE VOLUME system - A Computed Tomography X-Ray System intended to produce images of the head and body. Model
Hazard: Cooling Unit (CLU) expelled out of a CT gantry of an iCT system may result in injury
June 2, 2021Medical DevicesFDA
Philips North America Llc recalls IntelliVue MX40 Patient Monitor, Model Number: 865350 (US) 865352 (OUS) Software version number B.06.59 - Product Usage:
Hazard: Manufacturing failed to follow process steps resulting in the distribution of potentially defective devices
May 5, 2021Medical DevicesFDA
Philips North America Llc recalls Philips Xper Flex Cardio (FC2010, Procedure Room and FC2020, Bedside Monitor) devices - intended for use by professional
Hazard: Performance issues with the Xper Flex Cardio Physio Monitoring System include: potential delay of up to 10 seconds in displaying ECG, invasive blood pressure and other parameters; patient weight is rounded to the nearest whole kilogram;Xper IM software used with the Xper Flex Cardio Physio Monitoring System may periodically crash; No SpO2 numeric or plethysmography is displayed when SpO2 is connected o the Flex Cardio device; display of certain data from the FC2010 becomes frozen, i.e., waveforms cease sweeping and updating and the ECG, IBP, and respiration numeric values cease to update;e ECG, IBP, and respiration waveforms become flat lines and no audible alarms are produced for HR and IBP, upon start up, an unexpected non physiological ECG waveform, erratic heart rate numeric value, and non-physiological display of any other active waveforms may appear on the Boom Monitor
April 7, 2021Medical DevicesFDA
Philips North America, LLC recalls Philips SureSigns VM4, VM6 and VM8 - Product Usage: patient monitors-for monitoring, recording and alarming of multiple
Hazard: Fail to Comply with Chinese Standard YY1079-2008: Clauses 4.2.6 and 4.2.7.3. Range/accuracy of heart rate meter for pediatric mode-In pediatric mode, when the input signal rate is over 300 bpm, the indicated rate of the affected products will be lower than this upper limit
March 24, 2021Medical DevicesFDA
Philips North America Llc recalls Incisive CT Computed Tomography X-ray System Model: 728143
Hazard: Actuators in the Incisive gantry could result in the gantry failure to self-lock if both of the 2 sides actuators fail. In some cases in systems with affected components, the gantry may tilt forward slowly, in a worst case, reaching to the tilt limit angle +31degrees
February 3, 2021Medical DevicesFDA
Philips North America, LLC recalls TRx4841A 1.4 GHz IntelliVue Tele TRX, Model Number 862439; Includes System Codes 453564007261, 453564007271, 45356400
Hazard: The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
February 3, 2021Medical DevicesFDA
Philips North America, LLC recalls TRx4851A 2.4 GHz IntelliVue Tele TRX, Model Number 862231; Includes Service Numbers 453564052401, 453564052411, 45356405
Hazard: The ECG signal from patients being monitored using a Philips TRx4841A and TRx4851A Telemetry Transceiver may not be properly analyzed when it is used with a Philips Patient Information Center iX Release C.02.xx or C.03.01. If this occurs, the Information Center will not display a heart rate or generate, display or annunciate any heart rate or arrhythmia alarms.
January 20, 2021Medical DevicesFDA
Philips North America, LLC recalls Emission Computed Tomography System Image Process System - Product Usage: intended for general Nuclear Medicine imaging
Hazard: A software defect that has the potential to result in image misdiagnosis and incorrect treatment of a patient.
December 16, 2020Medical DevicesFDA
Philips North America Llc recalls Philips IntelliVue MP2 Patient Monitors (865040) - Product Usage: intended for monitoring and recording of, and to gener
Hazard: Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.
December 16, 2020Medical DevicesFDA
Philips North America Llc recalls Philips IntelliVue X2 Multi measurement Server: M3002A (865039) - Product Usage: intended for monitoring and recording o
Hazard: Speakers on IntelliVue X2 and IntelliVue MP2 may fail, identified by the display of a Speaker Malfunct. INOP or by the absence of audible sound, may result in delayed treatment of the patient.
November 4, 2020Medical DevicesFDA
Philips North America, LLC recalls Philips Sterilizable Defibrillator Paddles, Switchless Internal Paddles, Model Numbers M1741A, M1742A, M1743A, M1744A -
Hazard: The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.
November 4, 2020Medical DevicesFDA
Philips North America, LLC recalls Philips Sterilizable Defibrillator Paddles, Switched Internal Paddles, Model Numbers M4741A, M4742A, M4743A, M4744A - Pr
Hazard: The periodic Paddle Checks recommended in the Instructions for Use for Sterilizable Defibrillator Paddles may not detect one failure mode for the Switched Internal Defibrillator Paddles. Philips has created an addendum to the IFU, that includes a test for the switched paddles only, which will detect this failure mode.
October 7, 2020Medical DevicesFDA
Philips North America, LLC recalls ProxiDiagnost N90
Hazard: Unexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose.
August 26, 2020Medical DevicesFDA
Philips North America, LLC recalls Phillips Module Charger, Product #: 989803191031 - Product Usage: This Module Charger is used in the Expression MR400 MR
Hazard: A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
August 26, 2020Medical DevicesFDA
Philips North America, LLC recalls Phillips Charging Station, Product #: 989803191021 - Product Usage: This charging station is used in the Expression MR40
Hazard: A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
August 26, 2020Medical DevicesFDA
Philips North America, LLC recalls Phillips 3.7V Rechargeable Li-Polymer Battery, Product #: 989803191341 - Product Usage: This battery is used in the Expr
Hazard: A problem has been detected with the Philips LI-Polymer rechargeable battery when used with its approved charging station and module charger. When the battery is inserted into a charger bay, the LED indicatory lights solid red and the battery will not charge. This may result from poor electrical contact between the battery and charger.
August 19, 2020Medical DevicesFDA
Philips North America, LLC recalls Philips HeartStart MRx, Monitor/Defibrillators with model numbers M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M35
Hazard: The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user. These device behaviors could result in a delay in therapy or a failure to deliver the intended therapy.