PHILIPS MEDICAL SYSTEMS NEDERLAND B.V Recalls

209 recalls on record · February 23, 2022 to August 19, 2026 · Medical Devices

Recalls by year

22
23
24
25
26

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722028;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Number: 722027;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Number: 722026;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 OR Table; Model Number: 722023;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane; Model Number: 722013;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722012;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Number: 722011;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Number: 722010;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane; Model Number: 722008;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Number: 722006;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Number: 722005;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Number: 722003;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10F; Model Number: 722002;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
January 21, 2026Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Product Name: Allura Xper FD10C; Model number: 722001;
Hazard: Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
December 3, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion 7M20 systems with FlexArm ceiling-mounted system. Model Number: 722234;
Hazard: The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel. Expansion to add additional model.
October 29, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system
Hazard: Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Cardio Vascular-Allura Centron; Model Numbers: 722400; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper CV20; Model Numbers: 722031; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K16
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/20 OR Table; Model Numbers: 722039; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979,
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/20; Model Numbers: 722038; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859,
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/15 OR Table; Model Numbers: 722059; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979,
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/15; Model Numbers: 722058; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859,
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20/10; Model Numbers: 722029; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859,
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 OR Table; Model Numbers: (1) 722015, (2) 722023, (3) 722035; Associated 510(k)s: K102005, K130638, K1
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane OR Table; Model Numbers: (1) 722020, (2) 722025; Associated 510(k)s: K102005, K130638, K13084
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20 Biplane; Model Numbers: (1) 722008, (2) 722013; Associated 510(k)s: K102005, K130638, K130842, K13329
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD20; Model Numbers: (1) 722006, (2) 722012, (3) 722028; Associated 510(k)s: K102005, K130638, K130842, K1
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10 OR Table; Model Numbers: 722033; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K16
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Numbers: (1) 722005, (2) 722011, (3) 722027; Associated 510(k)s: K102005, K130638, K130842,
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
October 8, 2025Medical DevicesFDA
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K1
Hazard: The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.