June 5, 2019Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 17207 3/8 ECC PACK Material: 701049346
Hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
June 5, 2019Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 5103#CPS Tubing Pack Material: 701048690
Hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
June 5, 2019Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-T 8803#Circ Circuit Small Patient Material: 701048539
Hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
June 5, 2019Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 8013 ECC QUADROX iD Pack RF32 Material: 701047962
Hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
June 5, 2019Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Maquet Getinge-BEQ-TOP 15800 TUBING PACK Material: 3000065087
Hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
March 13, 2019Medical DevicesFDA
Maquet Cardiovascular, LLC recalls AXIUS Blower Mister Product Code/REF Number: CB-1000 The Axius Blower Mister is intended to clear an anastomotic sit
Hazard: Potential lack of carbon dioxide (CO2) flow that may result in a procedural delay during off-pump coronary artery bypass (OPCAB).
March 22, 2017Medical DevicesFDA
Maquet Cardiovascular, LLC recalls VASOVIEW HEMOPRO VH-3500 ENDOSCOPIC VESSEL HARVESTING SYSTEM Product Usage: The VASOVIEW HEMOPRO VH-3500 is designe
Hazard: Maquet has received several complaints involving the VASOVIEW HEMOPRO VH-3500 which involve the device exhibiting the failure mode of intermittent or no power.
November 5, 2014Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Acrobat V Vacuum Off-Pump System OM-9100S Acrobat SUV Vacuum Off-Pump System OM-9000S Axius Xpose 3 Access Devic
Hazard: An internal finding identified the presence of pinholes/cuts in some of the Tyvek pouches of the vacuum tubing sets included as an accessory to Acrobat V Vacuum Off-Pump System, Acrobat SUV Vacuum Off-Pump System, Axius XPOSE 3 Access Device and XPOSE 4 Access Device.
December 18, 2013Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Vasoview Hemopro Endoscopic Vessel Harvesting System Maquet Cardiovascular LLC 45 Barbour Pond Drive Wayne, NJ 07470
Hazard: Maquet has received an increase in the number of complaints for "self-activated/remains activated/overheats" for the Hemopro VH-30000, "Hemopro 1"
October 9, 2013Medical DevicesFDA
Maquet Cardiovascular, LLC recalls Ultima OPCAB System, Sterile, Rx Only, Product Usage: Intended for use during performance of minimally invasive car
Hazard: MAQUET has identified a potential issue with the ULTIMA OPCAB (part numbers OM-2003S & OM-2001D). Labeling for the 2 blade lots had been incorrectly applied. The standard blades from Lot #25052176 were incorrectly labeled as deep blades, and the 20 deep blades from Lot #25052200 were incorrectly labeled as standard blades.
April 24, 2013Medical DevicesFDA
Maquet Cardiovascular, LLC recalls MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp Syste
Hazard: It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl
January 16, 2013Medical DevicesFDA
Maquet Cardiovascular, LLC recalls CARDIOSAVE Intra-Aortic Balloon Pump An electromechanical system used to inflate and deflate intra-aortic balloons. I
Hazard: In the Cardiosave Intra-aortic Balloon Pump, there is a possibility that the coiled cord assembly, a prominent connector cord between the Cardiosave IABP housing and the dual display head which secured atop the IABP may fail as a result of mishandling.