Home / Brands / Intuitive Surgical
Intuitive Surgical Recalls
257 recalls on record · August 8, 2012 to July 22, 2026 · Medical Devices
Recalls by year
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.
February 12, 2020Medical DevicesFDA
Intuitive Surgical, Inc. recalls Da Vinci Xi/X EndoWrist Monopolar Curved scissors (MCS), model number 470179-19, UDI: (00) 886874112298
Hazard: This recall is being initiated because lots of EndoWrist Monopolar Curved Scissor instruments are not recognized by the integrated electrosurgical unit, which prohibits the activation of monopolar energy from the surgeon console.
February 12, 2020Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci SP Surgical System, SP1098, 00886874114605 Product Usage: used with an endoscope, surgical instruments, and ac
Hazard: The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
February 12, 2020Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci X Surgical System Surgical System, IS4200, 00886874114216 Product Usage: used with an endoscope, surgical inst
Hazard: The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
February 12, 2020Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci Xi Surgical System Surgical System, IS4000, 00886874114216 - Product Usage: used with an endoscope, surgical in
Hazard: The firm become aware of a potentially defective capacitors on circuit boards installed on certain da Vinci Xi, X and SP system Vision Side Carts (VSC) and/or Patient Side Carts (PSC). As a result, you may experience either vision loss or non-recoverable errors 307 or 319, preventing further use of the system.
February 5, 2020Medical DevicesFDA
Intuitive Surgical, Inc. recalls The da Vinci SP surgical system is made of three main components: the Surgeon Console (SC), the Vision Cart (VC), and th
Hazard: The firm became aware of a problem with the surgical system where the cannula mount button may be damaged when the user releases the button abruptly.
November 6, 2019Medical DevicesFDA
Intuitive Surgical, Inc. recalls EndoWrist Suction Irrigator, Single Use Instrument. Part Number: 480299-04. An accessory to the EndoWrist Surgical S
Hazard: The firm has identified the potential for a silicone particle to be generated within the Suction Irrigator device. If generated, a particle may be introduced into the patient with irrigation.
February 27, 2019Medical DevicesFDA
Intuitive Surgical, Inc. recalls daVinci Harmonic ACE Curved Shears
Hazard: During an internal inspection, Intuitive found that the sterile packaging of the da Vinci S/Si Harmonic ACE Inserts may become damaged during transit. The potential damage includes pinholes and /or small openings (less than a few millimeters) on the pouch leading to a potential breach in sterility.
October 3, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci Si/X/Xi Surgical System Redundant Medical Grade Power Supply (RMGPS) Units The da Vinci System consists of
Hazard: Certain Redundant Medical Grade Power Supply (RMGPS) units specific to Surgeon Console and Patient Side Carts were manufactured incorrectly and may fail prior or during use.
September 19, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci X/Xi EndoWrist 8mm Monopolar Curved Scissors Model No. 470179-14 Product Usage: EndoWrist Instruments, incl
Hazard: Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of the extension tube. These cracks may fracture the plastic surrounding the instrument shaft, and may cause fragments of the extension tube to fall, including potentially into the patient.
September 19, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci S/Si EndoWrist 8mm Monopolar Curved Scissors Model No. 420179-16 Product Usage: EndoWrist Instruments, incl
Hazard: Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of the extension tube. These cracks may fracture the plastic surrounding the instrument shaft, and may cause fragments of the extension tube to fall, including potentially into the patient.
August 22, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls The da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Hazard: An Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
July 18, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci Xi EndoWrist Stapler 45 Reload Green, Model 48445G-03, 12 reloads/box.
Hazard: Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.
July 18, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box.
Hazard: Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.
July 11, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci X Surgical System, Model IS4200. Product Usage: The Intuitive Surgical Endoscopic Instrument Control System
Hazard: Specific da Vinci X Surgical Systems (IS4200) were shipped with defective Arm Controller Joint (ACJ) boards that were found to have a high rate of failure due to a defective component.
January 31, 2018Medical DevicesFDA
Intuitive Surgical, Inc. recalls daVinci X EndoWrist(R) Stapler 45, REF 470298 Product Usage: The Intuitive Surgical EndoWrist Stapler 45, Stapler 45
Hazard: Specific da Vinci Xi EndoWrist Stapler 45 instruments are potentially impacted by a variation in the manufacturing process. In affected instruments, there is a possibility that the Stapler Release Kit (SRK) Wrench interface feature accessed through release hole 2 is manufactured to an incorrect dimension, which may prevent the SRK Wrench from being able to manually unclamp the instrument.
July 5, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci Xi Surgical System Product Usage: The da Vinci Xi Surgical System (IS4000) is designed to assist a surgeon
Hazard: da Vinci Xi(R) Surgical System Patient Side Carts (PSC) were shipped to the field with potentially under-torqued brake screws in a section of the system arms.
April 19, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B4
Hazard: Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Turkish SRK Genera
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Spanish SRK Genera
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Swedish SRK General
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Portuguese SRK Gen
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Korean SRK Gener
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Italian SRK Gene
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Japanese SRK Gen
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Greek SRK Genera
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi German SRK Gener
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Norwegian SRK Ge
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi French SRK Gener
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Finnish SRK Gene
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.
April 12, 2017Medical DevicesFDA
Intuitive Surgical, Inc. recalls STAPLER,IS4000; da Vinci Xi EndoWrist Stapler Release Kit (SRK); model number 470298. da Vinci Xi Dutch SRK Genera
Hazard: Intuitive Surgical has become aware and further investigation has determined that misinterpretation of instructions provided on the IS4000 Stapler Release Kit instruction card may result damage of the tool.