Integra LifeSciences Recalls

246 recalls on record · June 20, 2012 to March 12, 2025 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21
22
23
24
25

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

Also filed under

The recall feeds record what looks like the same company under more than one name:

February 18, 2015Medical DevicesFDA
Integra LifeSciences Corporation recalls The MAYFIELD¿ Ultra 360 Base Unit for head positioner, one unit per package. Product Usage: This product provide
Hazard: Through an investigation of complaints, Integra found that there is the possibility for the locking handle mechanism of the Base Units to fail during use.
December 24, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring Integra Licox Recon Tissue Oxygen Monitoring 1) LICOX¿ Complete Brain P
Hazard: Due to a complaint, it was determined that multiple lots of Licox catheters manufactured since October 29, 2013 contained a smart card which was programmed with an inaccurate lkN2 value.
December 24, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Ascension Orthopedics PyroSphere CMC & PyroSphere TMT, intended to replace the joint between the first metacarpal and th
Hazard: Integra LifeSciences has determined that a portion of some specific lots of size 10 PyroSphere CMC and PyroSphere TMT devices are non-radiopaque.
December 10, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Selector Kit Sterile single use device A bi-lumen, sterile single use accessory used for the Selector Integra U
Hazard: Some specific lots of Selector Tubing may potentially leak irrigation fluid during use if not placed on the Selector handpiece in a specific manner.
November 19, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Pt02 Monitor, Rx Only, Manufacturer: Integra LifeSciences (Ireland) Limited, IDA Business and Technology P
Hazard: Integra LifeSciences Corporation received 2 complaints that the USB port on the Licox Pt02 monitor does not consistently provide the user the ability to extract the Pt02 trend data according to the User's Manual for the device.
November 19, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Titan Reverse Shoulder System right and left head cutting templates, 142.5 degrees. The Reverse Shoulder Syst
Hazard: A single lot of left and a single lot of right Reverse Shoulder System cutting templates were manufactured incorrectly. Specifically, the threaded handle has been welded backwards on the template resulting in the cutting angle being the reverse of what it should be.
August 20, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra NewPort MIS System Model No. NWPIMP (NewPort Implant Tray w/ Components) NWP2INSTP (NewPort Instrum
Hazard: Integra LifeSciences is recalling the Integra NewPort MIS System because there is a possibility that components from the demonstration set may have been mixed with the NewPort MIS pedicle screw system components used in surgery.
July 23, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Camino 110-4 Intracranial Pressure (ICP) Monitoring Catheters, Cat. No's. 110-4B, 110-4G, 110-4H, 110-4L
Hazard: Integra LifeSciences is recalling the Camino Intracranial Pressure Monitoring Kits because of revised labelling to indicate they are unsafe in an MR environment.
July 9, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra¿ LED Battery Charger, Single Bay. Product Number: 90523 The LED Headlight System Battery Charger (90523) i
Hazard: LED battery chargers may prematurely fail and will not charge the LED battery as intended
July 9, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls MIS Rod 2, a component of the NewPort System. Model No: 60-0545 (45mm); 60-0550 (50mm); 60-0555 (55mm); 60-0560 (60mm
Hazard: Integra Lifesciences is recalling NewPort System MIS Rod, 2 because it may be up to 30 degrees offset from specification requirements causing misalignment with the inserter. This could potentially result in the lordosis of the rod not aligning with the handle when the inserter is used in accordance with the current NewPort MIS System Surgical Technique.
July 2, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Accell Evo3C Demineralized Bone Matrix, 5 cc, Model Number: 02-6000-050 , Lot Number: 132160. Product Usage:
Hazard: This lot of Accell Evo3C Demineralized Bone Matrix, 5cc was incorrectly packaged in boxes that describe the product as Accell Evo3 Demineralized Bone Matrix. Evo3C is contraindicated for treatment of vertebral compression and not indicated for use in the posterolateral spine.
June 11, 2014Medical DevicesFDA
Integra LifeSciences Corporation recalls Integra Newdeal SURFIX Fixed Angle Locking Screw 3.5mm x 22mm (catalog number 285322S) and 3.5mm x 30mm (catalog number
Hazard: New Deal has identified through a complaint report that some 3.5 mm Diameter x 22 mm Length Surfix screws in Lot no. F626 (Catalogue No. 285322S) have been packaged and labeled as 3.5mm Diameter x 30 mm Length Surfix screws (Catalogue No. 285330S).
June 11, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Test Kit REF BC10R Rx Only An accessory provided for use with the LCXO2 and AC31 PtO2 Monitors. It pro
Hazard: Integra LifeSciences has identified through a small number of complaints that some lot numbers of Licox Test Kits (Catalog Number BC10R) may provide erroneous "calibration test failure" message when they are used with the Licox PtO2 Monitor (LCSO2, or may provide an erroneous out of range result with the Licox CMP Brain Tissue Oxygen and Temperature Monitor (AC31).
May 7, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra CUSA Excel Disposable Wrench. Catalog Number C5601. The CUSA Excel+ Sterile 23 kHz Torque Wrench is an accessor
Hazard: Some 23 kHz Torque Wrenches were incorrectly assembled and contained a 36 kHz Hex Wrench Insert instead of the 23 kHz Hex Wrench Insert.
February 26, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Cusa Excel Rx Only CEM" Nosecone for CUSA¿ Excel System Product Usage: Intended for use CUSA Excel/CUSA
Hazard: The blue COAG button on the CUSA Excel CEM nosecone could stick in the 'activated' position.
February 19, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Malibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109; Malibu Polyaxial Screwdriver, Standard, Model Number 91-2
Hazard: Integra LifeSciences has identified through an investigation of complaints that there may be the potential for the shaft of the Malibu Polyaxial Drivers to break during use. None of the complaints reported a patient injury. However, some complaints did indicate the surgery was slightly prolonged while the additional driver provided in the kit was retrieved and used.
January 15, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Large Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive
Hazard: Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type packages were found prior to use but after the sterile package was open.
January 15, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Small Rickham -Style Reservoir Rx Only Sterile for Single Use Only Integra NeuroSciences 311 Enterprise Drive
Hazard: Integra became aware through an adverse complaint trend that Rickham Style Reservoirs with bent or broken connectors that were shipped in sleeve-type package were found prior to use but after the sterile package was open
January 8, 2014Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Camino Advanced Monitor with Waveform Display Model CAM01 Integra Neurosciences, 311 Enterprise Dr Plainsboro, N
Hazard: Integra became aware through a complaint from a single non-US customer that if the optional SYS/DIAS mode is used on CAM01 monitor that contains Mod 4 Firmware, the displayed values for the systolic and diastolic intracranial pressure (ICP) are reversed. But, if the CAM01 is used with an auxiliary bedside monitor, the bedside monitor displays these values correctly.
November 20, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Composite Series Skull Clamp, one clamp per tote. The MAYFIELD¿ Skull Clamp (A3059) is a cranial stabilization devic
Hazard: An investigation of an adverse trend of complaints for the Skull Clamp index locking knob not easily unlocking or disengaging and malfunctions of the 80 lbs. torque knob was initiated. None of the complaints resulted in a report of patient injury, adverse health consequence, or a delay / prolongation of the surgical procedure.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM3.ST_.EU Rx Only, Manufacturer GMS
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM3.S_.EU Rx Only, Manufacturer GMS-
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring 1 x Dual Lumen Introducer Kit REF IM2.S_.EU Rx Only, Manufacturer GMS
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring 1 x Dual Lumen Introducer Kit REF IM2_EU Rx Only, Manufacturer GMS- G
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM1.S Rx Only, Manufacturer GMS- G
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IM3.SR Rx Only, Manufacturer GMS-
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Inegra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Tunneling Probe Kit REF IT2 Rx Only, Manufacturer
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring 1 x Complete Brain Probe Kit REF IP2.P Rx Only, Manufacturer GMS- Ges
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring; 1 x Dual Lumen Introducer Kit REF IP2; Rx Only, Manufacturer GMS- Gesells
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.
October 30, 2013Medical DevicesFDA
Integra LifeSciences Corp. recalls Integra Licox Brain Tissue Oxygen Monitoring; 1 x Complete Brain Probe Kit REF IP1.P Rx Only, Manufacturer GMS- Gesell
Hazard: Integra LifeSciences Corporation has determined that there is a possibility that some lots of Licox CMP Brain Tissue Oxygen and Temperature Catheter Kits contain a hex wrench that does not loosen the set screw on the drill bit.