RecallWatchUSA

Home / Brands / Covidien

Covidien Recalls

493 recalls on record · June 27, 2012 to August 19, 2026 · Medical Devices

Recalls by year

12
13
14
15
16
17
18
19
20
21
22
23
24
25
26

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

Also filed under

The recall feeds record what looks like the same company under more than one name:

April 10, 2024Medical DevicesFDA
Covidien, LP recalls Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BT
Hazard: The potential for trocar seal disengagement when using mesh products incorrectly with the device.
April 10, 2024Medical DevicesFDA
Covidien, LP recalls Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTS
Hazard: The potential for trocar seal disengagement when using mesh products incorrectly with the device.
April 10, 2024Medical DevicesFDA
Covidien, LP recalls Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T12BT
Hazard: The potential for trocar seal disengagement when using mesh products incorrectly with the device.
April 10, 2024Medical DevicesFDA
Covidien, LP recalls Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTSNL
Hazard: The potential for trocar seal disengagement when using mesh products incorrectly with the device.
April 10, 2024Medical DevicesFDA
Covidien, LP recalls Covidien Auto Suture" Blunt Tip Trocar, Product Number OMS-T10BTNL
Hazard: The potential for trocar seal disengagement when using mesh products incorrectly with the device.
April 10, 2024Medical DevicesFDA
Covidien, LP recalls Covidien Auto Suture" Structural Balloon Trocar, Product Number OMS-T10SB
Hazard: The potential for trocar seal disengagement when using mesh products incorrectly with the device.
April 10, 2024Medical DevicesFDA
Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Hazard: Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
April 10, 2024Medical DevicesFDA
Covidien, LP recalls Signia" Small Diameter Curved Tip Intelligent Reload, Model SIGSDS30CTVT
Hazard: Under certain firing conditions, reloads were found to articulate in an uncontrolled manner potentially causing disruption to the staple line. The issue is related to components in impacted reloads that are not fully secure, which leads to uncontrolled articulation. The issue can occur using either the Endo GIA" Ultra Universal Stapler or Signia" Stapling System.
March 27, 2024Medical DevicesFDA
Covidien recalls Situate Laparotomy Sponge, RF and X-ray Detectable, REF: L1818-04P01C-1
Hazard: Packs including 5 laparotomy sponges may contain incorrect product with a blue loop, X-ray detectable feature, and ethylene oxide treatment; however, this incorrect product wasn't steam pre-treated prior to sterilization, and has no RF tag, which if non-sterile, may lead to infection, tissue trauma, and sepsis, and if in the scannable area, would lead a "CLEAR" display.
February 14, 2024Medical DevicesFDA
Covidien LP recalls AbsorbaTack Absorbable Fixation Device 30 Violet Absorbable Tacks 5 mm- -intended for fixation of prosthetic material t
Hazard: Packaging of the device may cause a low seal strength in the side seal of the foil pouch potentially resulting in a sterile/moisture barrier breach, compromising the product sterility
February 7, 2024Medical DevicesFDA
Covidien LP recalls Sofsilk Braided Silk sutures: CS-211 SOFSILK* 4-0 BLK 45CM CVF21 X36 GS-831 SOFSILK* 4-0 BLK 75CM V20 X36 S-176 SOFSILK
Hazard: Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
February 7, 2024Medical DevicesFDA
Covidien LP recalls Surgilon Braided Nylon sutures: 88861919-31 SURGILON* 4-0 BLK 7X75CMPCT 88861919-41 SURGILON* 3-0 BLK 7X75CMPCT 888
Hazard: Specific lots of sutures were sterilized with gamma doses that exceeded the range approved. The extra gamma levels may decrease the strength of the sutures over time, which could result in harms such as wound dehiscence and hemorrhage/blood loss/ bleeding at a critical level.
January 10, 2024Medical DevicesFDA
Covidien recalls Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD
Hazard: Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
November 8, 2023Medical DevicesFDA
Covidien recalls McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only
Hazard: Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
November 8, 2023Medical DevicesFDA
Covidien recalls McGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-017-000, Sterile EO, Rx Only
Hazard: Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
November 8, 2023Medical DevicesFDA
Covidien recalls Stolen McGRATH MAC Video Laryngoscopes, REF: 301-000-000
Hazard: Defective video laryngoscopes that did not pass quality tests established for the product, that were designated for destruction, were stolen, and subsequently distributed for sale. Defective device use may result in failed laryngoscopy and/or intubation.
October 11, 2023Medical DevicesFDA
Covidien recalls Puritan Bennett Cuff Pressure Manager, REF: 180-03
Hazard: Inadequate internal fixation of the power inlet port to the cuff pressure manager device causes the external power inlet to became dislodged within the device housing after several uses, which prevents the device from being charged, and could lead to treatment delay.
October 4, 2023Medical DevicesFDA
Covidien LP recalls Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm
Hazard: Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 88883
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Curved Extensions, Kit REF: 8888345637
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit REF: 8888
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, Kit" REF: 8888345629
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit REF: 8888
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, Kit" REF: 8888345611
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit REF: 88
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, Kit" REF: 8888345603
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions REF: 8888345520HP
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions REF: 8888345512HP
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637HP
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
August 23, 2023Medical DevicesFDA
Covidien LP recalls MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637
Hazard: Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus