266 recalls on record · December 1, 1969 to December 29, 2021 · Medical Devices, Vehicles

Recalls by year

69
71
13
14
15
16
17
18
19
20
21

Count of recall notices naming this brand, by the year the recall was announced. A quiet year is not proof of a safe year.

Also filed under

The recall feeds record what looks like the same company under more than one name:

December 29, 2021Medical DevicesFDA
Cook Inc. recalls Skinny Needle with Chiba Tip. RPN 090010, 090010-ET; GPN G14254, G15290. Rigid needle available with EchoTip to provi
Hazard: Specific lots may contain black debris on the exterior of the inner stylet and/or on the interior of the outer cannula. Local inflammatory reactions may occur if an affected product is used.
December 22, 2021Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer Ansel Modification Model Number: KCFW-7.0-18/38-45-RB-ANL0-HC Order Number/GPN: G44155 Int
Hazard: Flexor Check-Flo Introducer Ansel Modification mislabeled. Products packaged with a label stating 6FR may contain a 7FR device and vice versa
December 22, 2021Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer Ansel Modification Model Number: KCFW-6.0-18/38-45-RB-ANL0-HC Order Number/GPN: G44154 In
Hazard: Flexor Check-Flo Introducer Ansel Modification mislabeled. Products packaged with a label stating 6FR may contain a 7FR device and vice versa
November 24, 2021Medical DevicesFDA
Cook Inc. recalls Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter,
Hazard: Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.
November 24, 2021Medical DevicesFDA
Cook Inc. recalls Transseptal Needle, Trocar. RPN TSNC-18-71.0, TSNC-19-56.0. GPN G02364, G02365. Product consists of a transseptal ne
Hazard: Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.
November 10, 2021Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 The introduce
Hazard: Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.
November 10, 2021Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, Introducer Set, RPN KCFW-6.0-35-45-RB, GPN G09908 The introducer sheath incorporates a hyd
Hazard: Affected lots may be manufactured incorrectly with the radiopaque marker band located just below the Check-Flo proximal fitting instead of the distal tip. The issue may not be identified until the device is placed in the patient and viewed under fluoroscopy, leading to increased procedural time.
February 17, 2021Medical DevicesFDA
Cook Inc. recalls Cook Medical Kwart Retro-Inject Ureteral Stent Set 7.0Fr. 22-32cm -Used for retrograde injection during Extracorporeal S
Hazard: Manufactured with incorrect material, which may result in the stent being stiffer and potentially susceptible to degradation could potentially lead to breakage/fragmentation.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Tuohy-Borst Side-Arm Introducer (Shuttle Select), RPN KSAW-6.0-38-90-RB-SHTL-FLEX-HC, Order Number G31138 Fle
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Tuohy-Borst Side-Arm Introducer (Shuttle Select), RPN KSAW-5.0-38-90-RB-SHTL-FLEX-HC, Order Number G31133 Fle
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL1-HC, Order Number G56228 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL0-HC, Order Number G56223 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL1-HC, Order Number G49045 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-7.0-35-70-RB-HFANL1-HC, Order Number G56227 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-7.0-35-70-RB-HFANL0-HC, Order Number G56222 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-7.0-35-55-RB-HFANL1-HC, Order Number G49043 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-7.0-35-55-RB-HFANL0-HC, Order Number G56232 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-6.0-35-70-RB-HFANL1-HC, Order Number G56226 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-6.0-35-70-RB-HFANL0-HC, Order Number G56221 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-5.0-35-70-RB-HFANL1-HC, Order Number G56225 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-5.0-35-70-RB-HFANL0-HC, Order Number G56220 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-5.0-35-55-RB-HFANL1-HC, Order Number G56235 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-5.0-35-55-RB-HFANL0-HC, Order Number G56230 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-4.0-35-70-RB-HFANL0-HC, Order Number G56219 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL1-HC, Order Number G56234 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
December 30, 2020Medical DevicesFDA
Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL0-HC, Order Number G56229 Flexor Introducers and Guiding She
Hazard: There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.
November 11, 2020Medical DevicesFDA
Cook Inc. recalls Fixed Core Wire Guide, RPN TSCF-35-260-3, GPN G00517. The product is a curved wire guide with PTFE coating manufactured
Hazard: Affected product was dropped during manufacturing, which may result in a bent wire guide and/or the presence of loose foreign matter on the wire guide.