November 1, 2023Medical DevicesFDA
Carl Zeiss Suzhou Co., Ltd. recalls OPMI LUMERA 300, REF 6137
Hazard: Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.