September 25, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Model Number C21510, RSM Pneumatic Tubing Kit item #2, REF C11137;
Hazard: Beckman Coulter is recalling the DxI 9000 Access Immunoassay Analyzer due to undersized Pneumatics Tubing in the reagent storage module causing air leaks, compressor failures or sample processing errors. Four complaints were filed for this event that led to Beckman Coulter finding undersized tubing in stock and initiated corrective actions. Use of the affected device may result in errors leading to delay of results.
September 11, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative,
Hazard: Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.
September 4, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Hazard: When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.
August 21, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Discrete photometric chemistry analyzer for clinical use.
Hazard: Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.
August 21, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Hazard: The reason for this recall is Beckman Coulter has determined that when a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS); is running with the software versions 1.17 and below; and a sample tube is sent from the LAS to DxI 9000 Access Immunoassay Analyzer, an out of sequence message will be sent from the Analyzer to the LAS. This issue may result in multiple error messages (i.e., a sample pick and place (PnP) error or a no result obtained (NRT) flag may be displayed) and subsequently a risk of delays in processing/reporting patient test results and patient treatment.
July 31, 2024Medical DevicesFDA
Beckman Coulter Inc. recalls Beckman Coulter, REF: OSR61173, IGM 4x14 mL R1, 4x11 mL R2
Hazard: LIH Influence check settings for Lipemia are incorrect which may result patient samples being reported as false low results or cause a high result to report as normal
July 31, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Dxl 9000 Access Immunoassay Analyzer, Catalog #C11137.
Hazard: The analyzers were built with nonconforming USB cables made with brass contacts instead of specified gold contacts. Brass contacts may corrode over time, leading to loss of internal instrument communication and delayed patient results.
July 10, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, C11137; in-vitro diagnostic
Hazard: Beckman Coulter is recalling their DxI 9000 Access Immunoassay Analyzers, an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids, by correction. The reason for the recall is: Beckman Coulter has determined that the DxI 9000 Access Immunoassay analyzer accepts calibrator values scanned from multiple calibrator cards when an operator configures a new calibrator lot. The analyzer accepts the calibrator lot values from any scanned calibrator card, regardless of the assay test code or calibrator material lot number. The software is not validating the integrity of each calibrator barcode scanned to ensure they are all from the same calibrator card. This includes all software versions and all DxI 9000 instruments that are currently in the field. This issue does not impact calibrator material with 2D calibrator card barcodes. This may cause a failed calibration curve. The health risk to patients is dependent upon the analytes that could be impacted by the error, to include receiving delayed or erroneous results. The issue was identified by Beckman Coulter while conducting a field action on DxI 9000 Access Immunoassay Analyzers.
July 10, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter DxI 9000 Access Immunoassay Analyzer is a discrete photometric chemistry analyzer for clinical use. It
Hazard: The reason for the recall is the pneumatic tubing used in the analytical pick-and-place (PnP) assembly cable may fully or partially disconnect from the flex cable assembly fittings during normal operation. The DxI 9000 Access Immunoassay Analyzer may generate a System Event log message that indicates PnP errors. Any cancelled tests will have an associated SYS flag. The analyzer may enter the Red-system status because of this issue. The analyzer also may not eject RVs, where unbound conjugate could potentially splash up on the sidewall of the RV, not being properly washed away, resulting in an artificially higher signal (RLU). The analyzer may not detect this splashing and continue with normal operation, which can cause false low or high erroneous results. Consequently, the analyzer may cancel in-progress tests. These errors could potentially cause delay reporting patient test results and patient treatments. The degree of the potential health hazard is dependent on the analytes that are potentially delayed, especially those analytes that would be used in the acute setting when treatment decisions are immediate. In this particular recall the highest risk analyte is troponin. Health consequences could include permanent impairment or possibly death if medical treatment decisions are delayed due to a delay in obtaining patient results. The issue was identified by Beckman Coulter Inc. manufacturing staff at their location in Brea California on February 29, 2024, during in-process acceptance inspection activities.
April 3, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Access Substrate, REF 81906, For use with the Access Immunoassay Systems
Hazard: On 07 February 2024 Beckman Coulter made the decision to conduct a field action on Access Substrate (81906) lots: 234600, 337984, 338040, 338214, 338229, 338289, 338322, 338333, 338451, 338587, and 338538. Beckman Coulter has identified that Access Substrate bottles in the identified lots were released from manufacturing without labels. Missing label may cause some patient results reporting delay but will not cause any errorreous results. However, because labs are required to have redundancy such that delay would not significantly impact clinical management and because the defect is easily recognized by the user, the probability of harm is remote.
March 27, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Access Total T4, REF 33800, IVD, thyroxine reagent assay.
Hazard: Four lots of Access Total T4 reagent lots demonstrated imprecision outside of the 10% specification which may result in erroneous or delayed results.
February 14, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer running software version 1.16 or prior,
Hazard: There are potential performance issues found in the DxI 9000 Access Immunoassay Analyzer, including: 1) Access Ultrasensitive Insulin users who configure SI units (pmol/L) applies an incorrect conversion factor; 2) System provides numerical results that are below the lowest reportable result if laboratory chooses to report assay results in units of measurement other than the defined default units; 3) A result of 0 incorrectly reported when the system utilizes an expression that utilizes non-numerical symbols (e.g., > or <) to derive a calculated test as part of result reporting; 4) Reserve volume is enabled and the assay LIS code is not the same as the assay Test ID. These issues could potentially lead to erroneous results or delay reporting results for multiple analytes.
February 14, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Access Ultrasensitive Insulin only used on the DxI 9000 Access Immunoassay Analyzer, Catalog Number 33410
Hazard: There is an issue with the Access Ultrasensitive Insulin assay protocol file (APF) for use with the DxI 9000 Access Immunoassay Analyzer (APF 182 test definition version 6.7 and below). The Access Ultrasensitive Insulin APF applies an incorrect conversion factor when converting from system default units of ¿IU/mL to the International System of Units (SI units) of pmol/L.
January 31, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls iQ Waste Well Adapter, Part Number: 700-3393
Hazard: Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
January 31, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls REV 2 Flowcell Capsule Spare, Part Number: C04154
Hazard: Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
January 31, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls iQ200 Series Urine Microscopy Analyzers, Part Numbers: a) 700-3320, b) 700-3370, c) 700-3345, d) 700-3347, e)
Hazard: Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
January 31, 2024Medical DevicesFDA
Beckman Coulter, Inc. recalls Beckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947
Hazard: Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
October 18, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, REF C11137, Sofware Versions 1.15 and below; invitro diagnostic
Hazard: The firm has investigated and confirmed that when DxI 9000 is connected to Laboratory Information System (LIS, host), the sample barcode presented may be ignored and replaced with a sample ID that was previously in the same rack and position. This will cause a delay of result to the sample currently presented to the system (Sample B) and may result in an erroneous test result for the prior sample (Sample A), which gets associated with the results of sample B.
October 18, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls Access Thyroglobulin Calibrators
Hazard: It has been determined through an internal investigation that the alternate Access Thyroglobulin calibrator cards for use with the Thyg reagent packs distributed in April 2023 will cause a decrease of approximately 11% in patient results.
September 6, 2023Medical DevicesFDA
Beckman Coulter Inc. recalls Beckman Coulter, IVD, REF: OSR61171, IgA, 4 x 14 mL R1, 4 x 11 mL R2
Hazard: Lipemic interference failed to meet the performance specification listed within the IFU.
August 2, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls Ringed DxI Reaction Vessels (RVs)
Hazard: Beckman Coulter, Inc. is recalling Ringed DxI Reaction Vessels (RVs) as part of Unicel DxI Immunoassay System (600, 800 series). The reason for the recall is that the firm has identified 3 lots of RVs that failed an internally used service test to evaluate washing efficiency. Increased relative light units (RLU) suspected to be caused by the resulting rough interior surface may lead to artificially falsely elevated results for susceptible tests, in particular, sandwich assays including, but not limited to, hsTnI, Hybritech p2PSA, IPTH/IPTH IO.
July 19, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls COULTER DxH Diluent, REF 628017; diluent used for counting and sizing blood cells
Hazard: Specific diluent lot numbers showed that conductivity, osmolality, and pH were out of manufacturing specifications.
July 5, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls Access Hybritech p2PSA reagent, REF B03704 and REF A49752.
Hazard: Reagent lot numbers include an insufficient concentration of blocking reagent.
May 24, 2023Medical DevicesFDA
Beckman Coulter Inc. recalls Beckman Coulter Bicarbonate, REF: OS6637, 4x173 mL
Hazard: There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
May 24, 2023Medical DevicesFDA
Beckman Coulter Inc. recalls Beckman Coulter Bicarbonate, REF: OSR6237, 4x50 mL
Hazard: There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
May 24, 2023Medical DevicesFDA
Beckman Coulter Inc. recalls Beckman Coulter Bicarbonate, REF: OSR6137, 4x25 mL
Hazard: There is a potential for a decrease in calibration ODs for Bicarbonate reagent which could result in calibration failures and/or quality control failures.
May 24, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls UA WIN 10 Kit, Catalog Number C52900, in-vitro diagnostic device, urine analysis
Hazard: Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.
May 24, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls DxU 840m Iris Urine Microscopy Analyzer, Catalog Number C76947, in-vitro diagnostic device, urine analysis
Hazard: Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.
May 24, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls DxU 850m Iris Urine Microscopy Analyzer Catalog Number C49513, in-vitro diagnostic device, urine analysis
Hazard: Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have been investigated.
May 24, 2023Medical DevicesFDA
Beckman Coulter, Inc. recalls iQ200 Series Urine Microscopy Analyzer: a) iQ 200 Select, Part Numbers C10684 and 700-3345; b) iQ 200 Elite, Part Numbe
Hazard: Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have been investigated.