October 17, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls Baxter Clearlink System Buretrol Solution Set, 115" (2.9 m) with 150 mL Clearlink Burette (Ball-Valve Drip Chamber) (Non
Hazard: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
October 17, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls Baxter Clearlink System Buretrol Solution Set, 115" (2.9 m) with 150 mL Clearlink Burette (Ball-Valve Drip Chamber); a s
Hazard: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
October 17, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls Baxter Interlink System Buretrol Solution Set, 117" (3.0 m), with 150 mL Burette (Ball-Valve Drip Chamber) (Non-DEHP);
Hazard: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
October 17, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls Baxter Interlink System Buretrol Solution Set with 150 mL Burette (Ball-Valve Drip Chamber); a sterile Rx IV fluid pathw
Hazard: Baxter has determined that the ball-valve feature of the Buretrol Solution Sets may not function as expected, allowing air to flow past the valve and into the tubing at the completion of dose within the burette.
October 3, 2012DrugsFDA
Baxter Healthcare Corp. recalls Lactated Ringer's Injection USP, packaged in a) 250 mL AVIVA Container, product code 6E2322B, NDC 0338-6307-02; b) 500 m
Hazard: Chemical Contamination: The IV solutions were packaged in AVIVA containers which may contain trace levels of cumene hydroperoxide (CHP).
October 3, 2012DrugsFDA
Baxter Healthcare Corp. recalls 0.9% Sodium Chloride Injection USP, packaged in a) 250 mL AVIVA Container, product code 6E1322, NDC 0338-6304-02; b) 500
Hazard: Chemical Contamination: The IV solutions were packaged in AVIVA containers which may contain trace levels of cumene hydroperoxide (CHP).
October 3, 2012DrugsFDA
Baxter Healthcare Corp. recalls 5% Dextrose and 0.45% Sodium Chloride Injection USP, packaged in a 500 mL AVIVA Container, Rx only, Baxter Healthcare Co
Hazard: Chemical Contamination: The IV solutions were packaged in AVIVA containers which may contain trace levels of cumene hydroperoxide (CHP).
October 3, 2012DrugsFDA
Baxter Healthcare Corp. recalls 10% Dextrose Injection USP, packaged in a 500 mL AVIVA Container, Rx only, Baxter Healthcare Corporation, Deerfield, IL
Hazard: Chemical Contamination: The IV solutions were packaged in AVIVA containers which may contain trace levels of cumene hydroperoxide (CHP).
October 3, 2012DrugsFDA
Baxter Healthcare Corp. recalls 5% Dextrose Injection USP, packaged in a) 250 mL AVIVA Container, product code 6E0062, NDC 0338-6346-02; b) 500 mL AVIVA
Hazard: Chemical Contamination: The IV solutions were packaged in AVIVA containers which may contain trace levels of cumene hydroperoxide (CHP).
September 5, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls Mini-Infuser Model 300XL Multispeed Infusion Pump, an Rx battery operated syringe infusion pump with an active time scal
Hazard: Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
September 5, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls Mini-Infuser Model 150XL Single Speed Infusion Pump, an Rx battery operated syringe infusion pump; Product Usage:
Hazard: Should a battery with the minimum contact diameter (.51 inches) be placed on the battery spring (.53 inches), an electrical circuit will not be completed and the Mini-Infuser will not operate. The reason being that the Mini-Infuser Battery Spring has a flat contact that measures .53 inches ¿.01 inches on the inner diameter of the spring coil. The IEC 60086-2 and ANSI 08 standards specify a minimum
August 15, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls AUTOMIX 3+3 Compounder with Accusource Monitoring System, product 2M8287; automated nutrition compounders; Product U
Hazard: After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
August 15, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls AUTOMIX 3+3 Compounder System, product 2M8286 (115V) and 2M8286K (230V), automated nutrition compounders; Product Us
Hazard: After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
August 15, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls AUTOMIX Plus High Speed Compounder System, product 2M8075; automated nutrition compounders; Product Usage: The Auto
Hazard: After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
August 15, 2012Medical DevicesFDA
Baxter Healthcare Corp. recalls AUTOMIX High Speed Compounder System, product 2M8077; automated nutrition compounders; Product Usage: The Automix C
Hazard: After fluid ingress into the keypad of the AUTOMIX Compounder's (AUTOMIX) control module, the AUTOMIX may generate an incorrect device response to an AUTOMIX operator's key press. Also, the AUTOMIX system may experience an intermittent electrical failure, in which unexpected pump motor operations will occur. Also, the AUTOMIX system may intermittently experience an electrical failure where the compounder may stop compounding without an alarm and not be detected by the user.
June 20, 2012Medical DevicesFDA
Baxter Healthcare Corporation recalls Sigma Spectrum VOLUMETRIC INFUSION PUMP, with or without Master Drug Library, Part Number 35700 -- Common/Classification
Hazard: Recall expansion; the infusion pumps have the potential to fail causing inaccurate flow conditions during use. These conditions range from back flow to free flow, which could result in over-infusion. Sigma's evaluation of subsequent complaints involving failed bearings indicated that the loss or degradation of bearing lubricant to be a more significant contributor to bearing failure than initially believed. Based upon the results of the analysis of these additional failures, Sigma determined that any bearings prior to the latest, improved design may be at risk for failure. Sigma decided to expand the recall of the Spectrum Infusion Pumps to include units manufactured from 1/18/2005 through 11/01/2010.